ID551PN
Weighing Controller
8
2.3.2 2#Channel Scale
[2.1 ] Increment Size
0
(0.001),
1
(0.002),
2
(0.005),
3
(0.01),
4
(0.02),
5
(0.05),
6
(0.1),
7
(0.2),
8
(0.5),
9
(1),
10
(2),
11
(5),
12
(10),
13
(20),
14
(50),
15
(100)
[2.2 ] Capacity
Sets the maximum weighing of the scales, with default value 60
[2.3 ] Calibration Mode
2-P
:
two point calibration mode ,Only support
[2.4 ]
(Zero Point)&
[2.6 ]
(End
Point)
CALFr E
:
calibration without weighing added
[2.4 ] Zero Point Adjustment
See appendix 1.
[2.6 ] End-Point Calibration
See appendix 1.
[2.7 ] CalFree Block
[2.7.1 ]
Load Cell Capacity
Input the load cell capacity
[2.7.2 ]
Number of total load cells
Input total number of load cells in this scale
[2.7.3 ]
Number of working load cells
Input the number of working load cells in this scale
[2.7.4 ]
Load Cell Sensitivity
Input the sensitivity of load cell, in general, It is 2mV/V or 3mV/V
[2.7.5 ]
Sensitivity calibration
A test weight is loaded to calibrate the sensor sensitivity, and the method reference End Point
[2.8 ]
Filter
0
(
lowest
)
,
1
,
2
,
3
,
4
,
5
,
……18
,
19(
Highest
)
[2.9 ] Pushbutton Zero Range
0
–
Disable Pushbutton Zero
2
,
5
,
10
,
20
,
30
,
40
,
50
:
zero scale range is 2%
,
5%
,
10%
,
20%
,
30%
,
40%
,
50% of capacity
.
[2.10 ] Motion Check
[3.0d]
0
–
Disable motion check
0.1~9.9 –
Enable motion check