Appendix 2 Display Message & Error Code
Display Message &
Error Code
Description
CAL_2
Set Span Calibration
CAL_0
Set Zero Calibration
CAL_35
Calibration Fail, Check load cell cable or if scale is stability
CAL_D
Calibration Done
CALF-E
Factory CalFree Calibration data error, have to back to factory
for Factory Calibration again.
-EEE
Power Up Zero Fail for below zero too much
EEE
Power Up Zero Fail for above zero too much
EE - 10
E2PROM hardware error
EE - 11
E2PROM data error
EE – 100
Tare failed, Disable Tare
EE – 101
Tare failed for scale in motion
EE – 102
Tare failed out of tare range
EE - 105
Zero failed for Zero is disabled
EE - 106
Zero failed for scale in motion
EE - 107
Zero failed out of zero range
EE – 108
No keys allowed for keypad is locked
EE - 109
PLC option board start failed.
EE - 111
ADC Init failed, please re-power, or contact factory.
EE - no
The product key does not match, please contact the
manufacturer
--
Lock Display, change F10.F.3 to recover display weight
SEALEd
Seal lock, disable change weight parameters and weight
calibration,
Toggle switch to enable seal
Over capacity blank, weight is above scale capacity 9.9d
Under zero blank, weight is under zero 50d