ID551PN
Weighing Controller
16
1
Sign, Positive = 0, Negative = 1
2
Out of Range = 1 (Over capacity or Under Zero)
3
Motion = 1, Stable = 0
4
Always 1
5
Always 1
6
Always 0
7
EVEN/ODD Parity bit
SWC:
Status
C
Bit
0
Always 0
1
Always 0
2
Always 0
3
Always 0
4
Always 0
5
Always 1
6
Always 0
7
EVEN/ODD Parity bit
3.2 Continuous output A/D Code format
Output
Data
Start 1# A/D
Code
Separator 2# A/D
Code
Separator 3# A/D
Code
Separator
4# A/D
Code
END
Data
0x02 7 bytes comma
7 bytes comma
7 bytes comma
7 bytes
0x0D
Notice
:
A/D Code is 7 bits ASCII Code
,
less than 7 bits, fill 0, the Separator is comma (0x2C)
3.3 Demand Input / Output
Both Serial Port & Ethernet can be configured demand input/out put, below c ommands are
supported.
Demand Input
Demand character(2)
Description
NT
Tare Scale