ID551PN
Weighing Controller
10
[3.7 ] CalFree Block
[3.7.1 ]
Load Cell Capacity
Input the load cell capacity
[3.7.2 ]
Number of total load cells
Input total number of load cells in this scale
[3.7.3 ]
Number of working load cells
Input the number of working load cells in this scale
[3.7.4 ]
Load Cell Sensitivity
Input the sensitivity of load cell, in general, It is 2mV/V or 3mV/V
[3.7.5 ]
Sensitivity calibration
A test weight is loaded to calibrate the sensor sensitivity, and the method reference End Point
。
[3.8 ] Filter
0
(
lowest
)
,
1
,
2
,
3
,
4
,
5
,
……18
,
19(
Highest
)
[3.9 ] Pushbutton Zero Range
0
–
Disable Pushbutton Zero
2
,
5
,
10
,
20
,
30
,
40
,
50
:
zero scale range is 2%
,
5%
,
10%
,
20%
,
30%
,
40%
,
50% of capacity
[3.10 ] Motion Check [3.0d]
0
–
Disable motion check
0.1~9.9 –
Enable motion check
[3.11 ] Auto Zero Range
0
–
Disable Auto Zero
0.1~9.9 -
0.1d - 9.9d, Enable
When set to >0, the auto zero tracking is within
±
10d
[3.12 ] Auto Zero Speed
0
–
Disable Auto Zero
0.1~9.9 -
Enable Auto Zero, Speed is
(0.1~9.9)d/second
[3.13 ] Tare
0
– Disable Tare
1
–
Enable Tare
[3.14 ] Application Mode
0
– Weight mode
1
–
Dynamometry mode
[3.A ]
[3.A.1 ] :
View zero adjustment ADC code
[3.A.2 ] :
View span adjustment ADC code
[3.A.3 ] :
View span adjustment weight value