Technical characteristics
Product type
Medical thermometer in direct mode
Reference
Tucky-21
Medical device
Classe IIa
Normal condition of use
Ambient operating temperature: +15°C to 35°C
Technical accuracy cannot be guaranteed if the device is used outside the temperature and
humidity ranges indicated
Storage/transport conditions
Tucky-21W: Temperature: -15°C to 55°C ± 2°C. Humidity: 15% to 95% ± 5°C RH
ADH-21: Temperature: +10°C to 27°C ± 2°C. Humidity: 40% to 60% ± 5°C RH
Atmospheric pressure
Can operate at an atmospheric pressure which does not exceed 1 (760 à 1060 hPa)
Rate of humidity
Relative humidity range of 15% to 85% (without condensation)
Battery
Device with internal power supply: lithium battery. Capacity: 35mAh
Range of temperature displayed
+24°C to +44°C
Display resolution
0.1 °C
Accuracy
Accuracy for the range of temperatures displayed (maximum error in the laboratory): ±0.2°C
from 32°C to 42 °C then +/- 0.3°C outside this range
Technologie Bluetooth
4.2
Compatibility
Compatibility: from iOS 7.3 and Android 4.3
USB Cable
USB 2.0 Cable Type A Male to Type B Male
Charge voltage
5V
Protection index
IP22: Protected against solid objects larger than 12 mm.
Protected against vertically falling drops of water up to 15°C
Lifetime of Tucky
500 charging cycles
Lifetime of ADH-21 adhesives
5 years
Software classification
Class B
Warranty : please refer to the general sales terms, available on the website www.e-takescare.com
After-sales service
For any technical problems, please visit the www.e-takescare.com website
Not intended to be sterilised. Subject to modifications without notice.
IMMUNITY TESTING
TESTING LEVEL ACCORDING TO CEI 60601
LEVEL OF COMPLIANCE
Electrostatic discharge (ESD) CEI 61000-4-2
± 8 kV at contact
Compliant
± 2,4,8 and 15 kV in the air
RF radiated IEC 60601-4-3
10V/m de 80 MHz à 2,5 GHz
Compliant
Magnetic field at the network frequency IEC 60601-4-8
30 A/m
Compliant
ESSAIS ELECTROMAGNETIC EMISSION TESTING
COMPLIANCE
ENVIRONMENT
RF Emissions (CISPR 11)
Group 1
The RF emissions are very low and are not likely
to cause interference in nearby electronic equipment.
RF Emissions (CISPR 11)
Class B
Compliant
Harmonic current emissions
Not applicable
Battery power
Voltage fluctuation, flicker emissions
Not applicable
Battery power
Summary of Contents for TUCKY-21