Operation & Maintenance Manual

 

OMM1002 

 

 

 

Installation

 

1.

 

Check the instrument for any transit damage. Any damage to the by

-

pass loop could increase resistance to flow and therefore ren-

der calibration inaccurate.

 

2.

 

Re

-

check that the instrument is suitable for the chosen application.

 

3.

 

Check that the box includes (for complete instruments only):

 

 

a)   A Carrier of either Flanged or Grooved style.

 

 

b)   An Indicator by

-

pass loop, assembled for Horizontal pipelines & Left to Right flows.

 

 

c)   2 x Tube & Float Assemblies (1x assembled in Indicator housing).

 

 

d)   An impulse pipe for Vertically, Upwards flows.

 

 

e)   An impulse pipe for Vertically, Downwards flows.

 

4.

 

Check that the Platon Reference numbers on the Carrier and glass Tube are identical.

 

 

i.e. F1/** (see Table 2).

 

5.

 

Remove any packing materials, as necessary, from any component.

 

6.

 

For best accuracy ensure that the pipeline bores are smooth and that they match that of the Carrier.

 

7.

 

Select a location to ensure compliance with the Recommended Pipeline Configurations (see Fig 1 & Table 3).

 

Carrier Installation 

 

Flanged Style 

1.

 

The Carrier should be mounted inside the bolt circle of two flanges (see Specification). 

 

2.

 

Ensure the Carrier is precisely centred in relation to the pipeline & flanges.

 

3.

 

Ensure that the Gaskets (not supplies) either side of the Carrier has an Inside Diameter (I.D.) greater than or equal to the pipe nomi-
nal bore. Also ensure that when assembled no part of the gasket protrudes into the pipeline at any point.

 

 

Grooved Style 

 

1.

 

The Carrier should be mounted to the pipeline using two appropriately specified joints.

 

2.

 

Always lubricate gasket joints, so as to avoid pinching and leakage.

 

Both Carrier Styles 
 

1.

 

Ensure the Carrier flow label matches the flow direction of the pipeline.

 

2.

 

Ensure the process connections of the Carrier protrude from the pipeline in the correct direction to ensure:

 

3.

 

Vertical Indicator assembly.

 

4.

 

Readable Viewing angle.

 

 

Indicator Installation 

1.

 

Based on the pipeline orientation and direction of flow, assemble the Indicator by

-

pass loop, using the appropriate impulse pipe. 

Note the factory supplied set

-

up, is for Horizontal, Left to Right flows.

 

2.

 

Whatever the main pipeline configuration ensure that the Float moves freely in the glass Tube and that the indicator is accurately 
Vertical.

 

3.

 

Rotate the Indicator housing & glass tube, in relation to the rest of the Indicator by

-

pass loop, to obtain the best viewing angle.

 

Table 3

Recommended Minimum Upstream & Downstream Pipe Lengths (See Fig. 1)

 

Carrier 

 

Nominal Bore (mm) & Pipe Style Specification

 

50

 

80

 

100

 

150

 

200

 

BS1387

 

BS3600

 

L

 

250

 

400

 

500

 

750

 

1000

 

D1 Flanged

 

38.1

 

D2 Grooved

 

89.1  

 

98.1 

 

 

Min 5 x NB 

 

Page 3

 

Summary of Contents for SGUV MK.2

Page 1: ...Measurement Control LTD PLATON Issue B 07 2017 TT Electronics Roxspur Measurement Control 2 Downgate Drive Sheffield South Yorkshire S4 8BT UK Ph 44 0 114 244 2521 www ttelectronics com RoxspurSales ttelectronics com Every effort has been made during the preparation of this document to ensure the accuracy of statements and specifications However we do not accept liability for damage injury loss or...

Page 2: ...rier Installation Indicator Installation 3 3 3 Recommended Pipeline Configuration 4 Instrument Configuration Horizontal Pipelines 5 Instrument Configuration Vertical Pipelines 6 Operation 7 Maintenance 8 Glass Tube Cleaning or O Ring Glass Tube Replacement Filter Cleaning or O Ring Filter Element Replacement By pass Loop Air Removal 8 8 8 Spares 9 10 Returns Warranty 11 ...

Page 3: ...Indicator by pass loop Principles of Operation The SGUV Mk 2 is a combination of two simple measuring elements 1 Installed in the main pipeline is a sharp edged Orifice mounted in a Carrier ring This assembly produces a pressure drop proportional to the square root of the flow 2 Installed across the main Orifice Carrier assembly is an Indicator by pass loop containing an accurately calibrated Rest...

Page 4: ... Fire Security Products Services Part 5 Automatic Sprinkler Water Spray Deluge Systems Section 6 Direct Reading Flow Meters LPCB Ref NO Nominal Pipe Size mm Flow Range dm3 min Test Flow dm3 min Accuracy Test Flows dm3 min 088a 05 088a 10 088a 06 088a 11 088a 07 088a 12 088a 08 088a 13 088a 09 088a 14 50 80 100 150 200 150 750 300 2300 500 3500 900 8200 2500 12500 500 800 1300 1500 2200 2500 3500 4...

Page 5: ...anges 3 Ensure that the Gaskets not supplies either side of the Carrier has an Inside Diameter I D greater than or equal to the pipe nomi nal bore Also ensure that when assembled no part of the gasket protrudes into the pipeline at any point Grooved Style 1 The Carrier should be mounted to the pipeline using two appropriately specified joints 2 Always lubricate gasket joints so as to avoid pinchin...

Page 6: ...Operation Maintenance Manual OMM1002 4 Page 4 ...

Page 7: ...Operation Maintenance Manual OMM1002 5 Page 5 ...

Page 8: ...Operation Maintenance Manual OMM1002 6 Page 6 ...

Page 9: ... 3 Bleed all air from the Indicator by pass loop before taking any flow readings 4 Make sure that the Isolating Valves J are fully open before taking any flow readings 5 Read the top edge of the Float 6 When not in use the Isolation Valves should be closed and the Indicator by pass loop drained Once the unit has been fully drained all the Bleed Drain Valves should then be closed Page 7 ...

Page 10: ... end connectors by pulling noting which Insert came from which end connector To aid with their removal the plastic wedge can now be used as a sliding wedge 7 Remove 2x old O rings from each insert 8 Care should be taken that the new O rings C1 C2 are not damaged when assembling them to the inserts 9 Moistening the O rings with a suitable lubricant is advisable such that when re assembling to the e...

Page 11: ...he Air has been removed 4 Open the Isolation Valve Spares The instruments should not require maintenance to replace parts if the instrument is used in accordance with these instruc tions However spare parts are available to meet most needs as detailed below For Complete Instruments see Table 4 For Major Assemblies see Table 5 For Components Sub Assemblies see Table 6 Complete assembled Instruments...

Page 12: ...anged Grooved By Pass 50 LPCBC312 LPCB314 LGTF F1 5 LGTF F1 10 LPCB1 80 LPCBC512 LPCB514 LGTF F1 6 LGTF F1 11 100 LPCBC612 LPCB614 LGTF F1 7 LGTF F1 12 150 LPCBC812 LPCB814 LGTF F1 8 LGTF F1 13 200 LPCBC912 LPCB914 LGTF F1 9 LGTF F1 14 Table 5 Major Assemblies The Tube Float assemblies of a particular NB are universal between the 2 Carrier styles Two sets of codes have been included for simplicity...

Page 13: ...of being contaminated they will be returned to you On receipt of the appropriate documentation the goods will be examined and assessed in accordance with the terms of any Warranty agreement Repairs Prior to any repair work being carried out you will be informed of our findings and any charges that may be incurred Replacement Prior to the supply of any replacements you will be informed of our findi...

Reviews: