471C
17
Measurement
Operation
Mode Setting
(Set Code No.)
●
Option RS-232C, RS-485 output
Code
№
Function
Display
1
Setting range
Default
Value
80
Baud rate
BAUD.
4800, 9600, 19200bps
9600bps
81
Parity
PARIT
non(None), odd(Odd no.), even(Even no.) non
82
BCC switching
BCC
ON, OFF
OFF
83
Device number
RS.NO.
0 to 99
00
Setting method of code No.
This is the basic input method of function code.
Following the setting of the Code No., perform the function setting on the following pages.
Identification of the key color during
set up is as follows.
Black key: Valid key
White key: Invalid key
[
Common during the Setting mode
]
During setting mode, if the key is not
operated for about 5 minutes, the
operation returns into measurement
mode automatically. At this time, the
setting contents are not saved.
Again, if P
・
B key is pressed
for
more
than 1 second, the setting value
which may be under changing
process becomes invalid and returns
to measuring process.
MODE
▲
SET
SET
SET
MODE
Press 1 sec.
Confirm
Select the Code
№
,
▲
To function setting