P
ORTA
C
OUNT
®
Plus Model 8020 Technical Addendum
revised 7-27-2001
TSI Incorporated, St. Paul, Minnesota, USA
18
Copyright © 2001 all rights reserved
COMMAND
DESCRIPTION
PTPA0vv<cr>
Set Ambient Sample Purge Time
This command sets the ambient sample purge time for exercises 1 through 12. vv is
the ambient sample purge time in seconds and must be in the range from
04
to
25
.
For example, to set the ambient sample purge time to 8 seconds, send the command:
PTPA008<cr>
The PORTACOUNT Plus will respond with an echo of the same command:
PTPA008<cr><lf>
PPxxvvvvv<c
r>
Set Fit Factor Pass Level
The PORTACOUNT Plus stores a set of 12 fit factor pass levels in memory. The
operator is allowed to scroll through and select one of
these values as the current fit factor pass level using the PORTACOUNT Plus
Keypad.
This command allows any of the 12 fit factor pass levels stored in the
PORTACOUNT Plus to be permanently changed.
xx
is the memory location from
01 to 12.
vvvvv
is the new fit factor pass level between
00000
and
64000
. All fit
factor pass levels must have 5 characters. Use leading zeros when necessary, such as
000350
for 350 and
02000
for 2000.
For example, to change the 3rd of 12 fit factor pass levels to 350, send the command:
PP0300350<cr>
The PORTACOUNT Plus will respond with an echo of the same command:
PP0300350<cr><lf>