P
ORTA
C
OUNT
®
Plus Model 8020 Technical Addendum
revised 7-27-2001
TSI Incorporated, St. Paul, Minnesota, USA
16
Copyright © 2001 all rights reserved
COMMAND
DESCRIPTION
S<cr>
(continued)
(continued from previous page)
13 lines will be transmitted, one for each exercise. x
x
is the exercise number (
01
to
13
) and vv is the mask sample time in seconds. Mask sample times will be in the
range from
10
to
99
seconds. The 13
th
line is the mask sample time (60 seconds)
used when the number of exercises is set to zero (8010 Mode). The 13
th
mask sample
time cannot be changed.
Fit Factor Pass Levels
SP xxvvvvv<cr><lf>
12 lines will be transmitted, one for each of the fit factor pass levels stored in
memory. The PORTACOUNT Plus stores 12 fit factor pass levels in memory and
allows an operator to scroll through the list using the Keypad.
Xx
is the memory
location between
01
and
12
. Vvvvv is the fit factor pass level contained in that
memory location. The maximum value allowed for the fit factor pass level is
64000
.
There is one space after the
SP
.
Serial Number
The PORTACOUNT Plus serial number is transmitted in the format:
SS vvvvv<cr><lf>
vvvvv
is the serial number. There are three spaces after the
SS
.
Run Time Since Last Serviced at the Factory
The run time since the last time factory service was provided is transmitted in the
format:
SR vvvvv<cr><lf>
vvvvv
is the run time in 10 minute increments. There are three spaces after the
SR
.
For example:
SR 05370<cr><lf>
indicates that the unit has run for 53700 minutes or 895 hours since last serviced.
The timer is reset at the factory when service is performed.
Date Last Serviced
The date when the PORTACOUNT Plus was last serviced at the factory is
transmitted in the format:
SD 0MMYY<cr><lf>
Where
MM
is the month number (
01
to
12
) and
YY
is the last 2 digits of the year (
91
= 1991,
93
= 1993,
00
= 2000 etc.). The character before the month number (
MM
)
is always a zero. There are three spaces after the
SD
.