5 MAINTENANCE  & TROUBLE-ShOOTINg

5.1 BATTERy MAINTENANCE

Battery pack operating time is 10 hours continuous 
usage.  The batteries can be recharged at any time 
using the mains charger supplied.  Connect the battery 
charger to the Flowmaster battery charger socket.

 It is 

better to top up charge the batteries at regular 

intervals than to allow them to discharge fully. 

The batteries can be left on continuous charge when 
the Flowmaster is not in use. The batteries have an 
operating life of ten years and will need to be replaced 
thereafter.  The operating time will reduce as the 
batteries age.

EFFECT OF CONTINUOUS METER USE ON 

BATTERy LIFE

The type of battery in use on the Flowmaster 250 is 
NiMH type. This battery type is designed for regular but 

intermittent use. It is not designed for deep discharge 
such as would happen were the Flowmaster repeatedly 
left powered on for eight to ten hours per day. In 
these circumstances, the battery life would be reduced 
to ten months. Battery replacement under these 
circumstances is not covered under warranty.
Should your working procedures require that the 
Flowmaster is powered on for up to eight hours per 
day, please contact TSI Flowmeters Ltd and request an 
alternative battery technology.

TEST FOR RE-ChARgE REQUIREMENT

The  flow  measurement  will  become  erratic  if  the 
battery voltage is too low. The readout may indicate 
zero even when water is flowing or the flow reading 
may vary considerably even though the flow stream is 
steady. Use a voltmeter to check the voltage level of 
the battery pack. If the voltage is less than 10V DC 
then immediate recharging is necessary. If the voltage 
is less than 12V DC then you will be able to use the 
Flowmaster for some time, but immediate top-up 
charging is recommended.

ChARgINg ThE  BATTERIES

There are two Flowmaster battery charging sockets. 
One socket is located in the power panel located at 
the side of the Flowmaster. The other socket is located 
in the red casting on the other side of the Flowmaster. 
Connect the battery charger to one of the two sockets. 
Plug the charger into mains. Please refer to the battery 
charger instruction manual, document no. 5110A.

5.1.1

5.1.2

5.1.3

The  Flowmaster is practically maintenance free. 

 

Periodically charge the batteries. TSI Flowmeters 
recommend charging the batteries after each use and 
once per month when the meter is not being used. 
Clean the electrodes (especially if the meter has been 
used with oily water) to ensure trouble free operation 
of the flowmaster.
Please avoid dropping the Flowmaster.  Please use 

your hands to operate the switches and avoid use of 
screwdrivers.  Please keep abrasive chemicals away from 
the  exterior  of  the  Flowmaster  and  wash  the  inner 
tube with clean water after flowing foam or seawater.
The flow meter control electronics are in a waterproof 
compartment located beneath the display gauge. Please 

do not open this compartment.  Breaking the seal will 
destroy the water-proof feature of this compartment 
and void warranty.  Subsequent water ingress will 
damage the control electronics and lead to loss of flow 
measurement function.

5.2 FLOWMASTER MAINTENANCE

12

Fig. 5.1(a) Flowmaster battery charger socket (power 

panel)

Fig. 5.2 Location of electrodes

Electrodes

Fig. 5.1(b) Flowmaster battery charger socket (casting)

CLEANINg SENSOR ELECTROdES

The Flowmaster’s flow gauge response is reduced if the 
electrodes are covered with a film of oil, dirt or rust. 
This can result in measurement error which is usually in 
the form of a lower than expected flow reading.
Clean the electrodes by brushing the inside of the pipe 
with a soft cloth or soft brush dipped in water.  Use 
of  a  household  handwash  type  dishwashing  fluid  is 
acceptable.  If necessary, an alcohol or methanol based 
cleaning-agent may be used.
The electrodes should be cleaned periodically.  The 
frequency of cleaning depends on the type of water 
being used.  If clean tap water is used, then once a year 
is sufficient; if river water is used, then we recommend 
that they be cleaned every 6 months.  If oily water has 
been used, then they should be cleaned immediately 
after use by running clean tap water.

5.2.1

Summary of Contents for EMF-300E

Page 1: ...ie Unit 1 Portlaoise Enterprise Centre Clonminam Business Park Portlaoise Co Laois R32VK07 Ireland t 353 0 57 866 3852 e info tsi ie The Flowmaster 250 measuresthe flow rate andthe pressure at every w...

Page 2: ...Centre Clonminam Business Park Portlaoise Co Laois R32VK07 Ireland t 353 0 57 866 3852 e info tsi ie Version 1 7 June 2017 OPERATION MANUAL AND MAINTENANCE MANUAL Flowmaster 250 and Flowmaster 400 in...

Page 3: ...E SHOOTING 6 NiMH BATTERY CHARGER 6 2 HOW TO USE THE CHARGER 6 1 MASCOT CHARGE DIAGRAM 7 ACCESSORIES 7 2 TRANSPORT STORAGE CASE 7 1 BALLVALVE WITH BS INST M F FITTINGS 7 4 FLOW MONITORING SYSTEM 7 3 A...

Page 4: ...in components of the Flowmaster 1 4 1 1 INTRODUCTION The Flowmaster is now used by most UK and Irish fire services and has been exported to 50 countries The Flowmaster meets the requirements of the JO...

Page 5: ...in the sensor 9 Flashes on and off to indicate there may be dirt on the electrodes Ignore this signal when there is no water in the sensor 1 8 9 2 3 5 6 4 7 EXPLANATION OF USER PANEL 1 3 APPLICATIONS...

Page 6: ...esistant ADC6 12 JIS aluminium anodised Pressure rating 49 bar Digital LCD display gauge 4 digits 17 6mm high digits Analog bar graph Internal backlighting Type Bourdon Tube Range 0 10 bar 1 0 7 0 15...

Page 7: ...display in flow measurement mode is as shown below Fig 3 3 Flow meter display when ready for flow measurement and there is no water in the pipe Note the flashing symbols on the left side of the flowm...

Page 8: ...e second and then releasing 4 Figure displayed will automatically re set to zero The flow meter has a built in totalising function The volume of water that has passed through the flow meter is totaled...

Page 9: ...for static and residual pressure measurements and for flow measurements is supplied on the following pages This information is provided for guidance only It is preferential that the Flowmaster user fo...

Page 10: ...and this is very dangerous Fig 4 6 Caution Never allow the Flowmaster to be used in an open ended position It is essential to secure the Flowmaster so that it cannot move about freely 10 4 2 FLOW RES...

Page 11: ...water begins to drain from the ball valve Close the ball valve 5 When reading is taken close hydrant fully 4 Read static pressure in bar 6 Open the ball valve to exhaust pressure from the meter and s...

Page 12: ...in the red casting on the other side of the Flowmaster Connect the battery charger to one of the two sockets Plug the charger into mains Please refer to the battery charger instruction manual document...

Page 13: ...speaking if battery voltage falls below 6 volts DC it will not recover 5 3 TROUBLE SHOOTING INCOMPLETE SELF TEST SEQUENCE Display goes through the Self Test Sequence until this display 5 3 1 SCALE AD...

Page 14: ...Version 1 7 June 2017 TSI Flowmeters Ltd 6 NiMH BATTERY CHARGER 6 1 MASCOT CHARGE DIAGRAM 6 2 HOWTO USETHE CHARGER 14...

Page 15: ...S 7 1 BALLVALVE WITH BS INST M F FITTINGS The ball valve facilitates speedy and safe static pressure readings and creates back pressure for taking flow and residual pressure measurements 7 2 TRANSPORT...

Page 16: ...be quickly and easily generated using the accompanying Flow Monitoring Software Up to 6 hours of data can be captured at 1 second intervals Flowrates can be set that automatically start and stop the d...

Page 17: ...gations hereunder If Warrantor elects to comply with this warranty by means of such refund as a condition precedent to such compliance the claimant shall return such equipment toWarrantor free and cle...

Page 18: ...lowmaster Model 250 to which this documentation relates is in conformity with the provisions of the following directives 2001 95 EC General Product Safety 2006 95 EC Low Voltage Directive LVD 2004 108...

Page 19: ...Flowmeter Calibrated Type of Meter TSI Model EMF 300A Magflow Serial No Date of Calibration Method of Test Comparison with Master Flowmeter The zero offset was adjusted to read 0 lpm Since this is the...

Reviews: