Display does not reach this display of the Self-Test
Sequence -
Possible Cause:
The Sensor part of the device and the Display part of
the device are not in contact.
Action:
Check all cables are secure. If still not connected –
please contact your local representative
* This condition may indicate that water ingress to the
electronics compartment has damaged the control
electronics. Please contact your local representative.
ERRATIC dISPLAy
5.3.2
Possible Cause:
The battery voltage is too low.
Action:
Charge the batteries.
REAdINg NOT dISPLAyINg
The power-on Self-test Sequence is okay, but there is
no reading when water is flowing.
5.3.3
Possible Cause:
1) Batteries voltage is too low
2) No earth wire to battery
3) Unit is faulty
Action:
1) Charge or replace batteries
2) Replace earth wire between battery negative
terminal and red casting
3) Return unit to Manufacturer
REChARgINg ISSUES
The unit will not charge or there are humming noises
while trying to recharge.
5.3.4
Possible Cause:
Faulty battery charger or batteries are over-discharged
and cannot recover.
Action:
Contact local representative for replacement charger
or batteries. Generally speaking, if battery voltage falls
below 6 volts DC, it will not recover.
5.3 TROUBLE-ShOOTINg
INCOMPLETE SELF-TEST SEQUENCE
Display goes through the Self-Test Sequence until this
display -
5.3.1
SCALE AdJUSTMENT
The Flowmaster is calibrated and the scale adjusted
at the factory. Accurate measurements can be made
immediately upon receipt of the unit. It is recommended
that all measuring instruments are calibrated annually. If,
perchance, the display does not seem to be accurate,
then please check that the electrodes are clean.
5.2.2
5.4 CALIBRATION OF yOUR FLOWMASTER
Every TSI flow meter is tested and calibrated prior to
shipping. A calibration label is attached to the top lid of
the meter (Figures 5.3 and 5.4). The calibration date and
the date on which the meter is due its next calibration
are written on the label for your convenience.
Fig. 5.3 Calibration label on lid of Flowmaster
Fig. 5.4 Calibration label showing calibration date and
re-calibration date
The master meter at TSI is calibrated annually at the
Irish National Test Centre. This ensures the readings
obtained by the Test Equipment referred to on
certificates issued by TSI Flowmeters Ltd are traceable
to National and International standards
Annual calibration of the Flowmaster is recommended,
particularly if the meter is used on a regular basis.
Please return your meter to TSI Flowmeters Ltd for
calibration.