Chapter 6
Troubleshooting
The table below lists the symptoms, possible causes, and
recommended solutions for common problems encountered with
the DustTrak
™ monitor.
Symptom
Possible Cause
Corrective Action
Erratic reading.
Leak
Check connections for leaks
Dirty inlet port and/or
sample tube
Clean inlet port. Clean or
replace tubing
Internal filter(s) not
installed properly
(leaking)
Inspect internal filter wells to
make certain the filters and
O-rings are seated properly.
Replace internal filters if
necessary
DustTrak
reading
negative
concentrations
Zero Drift
Increase autozero
frequency
Analog output
does not work
Cable/connector not
correctly installed
Make sure cable connector
is fully seated
Output wired with
reverse polarity
Make sure analog out (+)
and analog ground (-) are
wired correctly to data-
logger
Alarm output
does not work
Alarm function not
turned on
Turn the alarm function on
in TrakPro
™ software
Alarm does not
turn on correctly
Alarm setting
incorrect
Check the alarm settings in
TrakPro
™ software
Alarm output wired
with reverse polarity
Alarm wires are polarized.
Voltage input must be wired
to alarm input (+)
Summary of Contents for DustTrak 8540
Page 2: ......
Page 14: ...4 Chapter 1 This page intentionally left blank ...
Page 20: ...Chapter 2 10 This page intentionally left blank ...
Page 62: ...Chapter 5 52 This page intentionally left blank ...
Page 68: ...58 Appendix B This page intentionally left blank ...
Page 72: ...62 Appendix C This page intentionally left blank ...
Page 77: ......