Chapter 5
44
5. Replace cover and return to
service (Figure 23).
6.
It is important to reset the
instruments filter change
date and cumulative filter
mass concentration after
replacing the Sheath air
filter.
Resetting the filter
concentration may clear the
filter error condition shown in
the Data tab of DustTrak
™
Checkout Software.
Reset
the date and concentration
by performing the following:
Figure 23: Replace cover
a. Power the instrument.
b. Open the DustTrak
™ Checkout software.
c. Connect the provided USB cable to the instrument and your
personal computer.
d. Click
Connect.
e. Click
Reset Filter Date and Concentrations.