![TSI Instruments DustTrak 8540 Operation And Service Manual Download Page 12](http://html1.mh-extra.com/html/tsi-instruments/dusttrak-8540/dusttrak-8540_operation-and-service-manual_1178212012.webp)
2
Chapter 1
Models 8540-M, 8542-M, and 8543-M MCERTS
Certified Monitors
DustTrak™ Environmental Monitor Models 854201-M1, 854001-M1 or
854301-M1, have been designed to meet
the UK’s MCERTS
(Monitoring Certification Scheme) for indicative instruments.
The
following must be installed for these models to meet the
MCERTS Indicative Certification:
854201-M1: MCERTS PM10 Indicative Certification
❑
8542-M Photometer
❑
854038 Environmental Enclosure
❑
854020 PM10 Impactor
❑
854041 Heated Inlet
❑
PCF set at 0.53 (preset at the factory). It is recommended that the
instrument is run with a Photometric Correction Factor (PCF) of
0.53. However, the application of a site specific correction factor
may be used if a significant decrease in the expanded uncertainty
can be demonstrated.
854001-M1: MCERTS PM2.5 Indicative Certification
❑
8540-M Photometer
❑
854038 Environmental Enclosure
❑
854021 PM2.5 Impactor
❑
854041 Heated Inlet
❑
PCF set at 0.33 (preset at the factory). It is recommended that the
instrument is run with a Photometric Correction Factor (PCF) of
0.33. However, the application of a site specific correction factor
may be used if a significant decrease in the expanded uncertainty
can be demonstrated.
Summary of Contents for DustTrak 8540
Page 2: ......
Page 14: ...4 Chapter 1 This page intentionally left blank ...
Page 20: ...Chapter 2 10 This page intentionally left blank ...
Page 62: ...Chapter 5 52 This page intentionally left blank ...
Page 68: ...58 Appendix B This page intentionally left blank ...
Page 72: ...62 Appendix C This page intentionally left blank ...
Page 77: ......