Operation
33
Use Start Date
Use Start Date,
option to use programmed
start date or by pass programmed start
date.
Use Start Time
Use Start Time, option to use programmed
start time or bypass programmed start
time.
In Log mode, data will be stored to a file named “
LogName_XYZ
” where
LogName
is the user entered log name and
XYZ
is an incrementing integer.
Taking Mass Concentration Measurements
Measurements are started and controlled from the main screen.
Prior to starting a measurement the instrument should be zeroed from the
Setup
screen and the run mode should be configured and selected from the
RunMode
screen.
When the instrument is on, but not taking any mass measurements the start
button will be green and instruments pump will not be running. To start
taking a measurement, press the green start button.
While taking a measurement the screen will display the current measured
mass concentration. The various regions of the screen are shown below.