Chapter 2
12
Dorr-Oliver Cyclone
A Dorr-Oliver cyclone is shipped with the instrument. The Dorr-Oliver
cyclone removes particles over 4.0 µm in size. The Dorr-Oliver cyclone is
attached to the instrument by sliding the cyclone clip over the protruding
catch. The tube from the Dorr-Oliver cyclone needs to be routed to the inlet
of the instrument.
Figure 2-5: Installing Door-Oliver Cyclone
Inlet attachments (impactors or inlet cap) should
not
be used when using the
Dorr-Oliver Cyclone.
The instrument flow rate must be changed to
1.7 L/min when using the Dorr-Oliver Cyclone in order to achieve a
4 µm (respirable) cut-point.
See the
Flow Cal
instructions in the
Operations chapter for instructions on how to change the instruments
flow rate.
Instrument Setup
The D
UST
T
RAK
™
II monitor can be connected to a computer to download
data and upload sampling programs.
Connecting to the Computer
Connect the USB host port of a Microsoft Windows®-based computer to the
USB device port on the side of the D
UST
T
RAK
™
monitor.
®
Windows is a registered trademark of Microsoft Corporation.