39
4 Maintenance
4.1 Recharging the Batteries
(as required)
The batteries can be charged internally by connecting the AC adapter or they can be
charged externally with the optional battery charger. The fastest charging method is to
charge the battery in the instrument with the instrument turned off. The Certifier FA Plus
uses SBL-160, Lithium Ion Batteries, which are available from many battery supply
houses or TSI. Charging time is approximately 4 to 6 hours.
4.2 Replacing the Oxygen Sensor
The oxygen sensor will function for one or more years of normal operation if use begins
before the expiration date. Replace the oxygen sensor every year or two of normal use,
or if the sensor cannot be calibrated or sensor readings are erratic.
4.3 Cleaning
(as required)
Table 5 summarizes recommended cleaning methods for Certifier
®
FA Test System
components.
Table 5. Cleaning Recommendations
Component
Cleaning
Interface Module
Flow modules
Carrying cases
Oxygen sensor
Oxygen sensor cables
Clean exterior as required with a clean cloth and
isopropyl alcohol, hydrogen peroxide (3%), or ammonia
(15%).
Tee
Adapters
Steam autoclave after contact with any non-sterile
breathing circuit components, and discard if any
damage is visible.
Single use filters
(High and Low
Flow Modules)
Discard after contact with any non-sterile breathing
circuit components or if damage is visible.