18
Certifier
®
FA Plus Test System
2.3.1 Measurement Selection
To add a parameter to the
Parameter Screen, touch the
parameter and then touch the
right direction arrow. The
available parameters change
depending on what module is
attached.
To remove a parameter from
the Parameter Screen, touch
the parameter and then touch
the left direction arrow.
To move a parameter higher
on the Parameter Screen,
touch the parameter and then
use the up and down direction
arrows.
Figure 19. Measurement Selection Screen
To change the gas calibration used, touch the gas dropdown list and select the
desired gas.
Note:
If the gas selection Air/O
2
Mixture is selected but no oxygen sensor is
plugged into the Model 4081 high flow module, then the high flow module
will use the internal air calibration only. No corrections for oxygen
concentration will be made if the oxygen sensor is not plugged into the
flow module.