![TSI Instruments CERTIFIER FA PLUS Operator'S Manual Download Page 36](http://html1.mh-extra.com/html/tsi-instruments/certifier-fa-plus/certifier-fa-plus_operators-manual_1178281036.webp)
30
Certifier
®
FA Plus Test System
Lock
Configuration
Disables the saving of configurations. The user will be able to load
existing configurations and change the current configuration, but not
save it.
Advanced
Features
for a complete list of advanced features which
can be accessed through this menu.
About Device
Displays information about the Certifier Interface Module and any
connected flow module.
Select Mass
Storage Target
When the Certifier test system is connected to a host computer via
USB interface, it appears to the host as a "mass storage" device,
similar to a "memory stick" or “thumb drive”. This function allows
you to select whether the host can access the Certifier's Internal
Memory or an SD Flash memory card. The Certifier test system
supports SD flash cards up to 1 GB in size. Restart the Certifier
whenever a change if made when selecting between internal flash
RAM and SD flash card.
2.3.10.1 Advanced Features
Figure 29. Advanced Function Menu
2.3.10.1.1 Configuration Import/Export
This feature enables you to both download and upload configuration files to and from
the SD card. This is convenient when multiple Certifier FA+ units are in use where the
configuration files from one unit can be transferred to multiple Certifier units. This
eliminates the need to generate configuration files for each instrument.