![TSI Instruments AEROTRAK 9306 Operation Manual Download Page 77](http://html1.mh-extra.com/html/tsi-instruments/aerotrak-9306/aerotrak-9306_operation-manual_1178118077.webp)
A-1
A P P E N D I X A
S p e c i f i c a t i o n s
All specifications meet or exceed ISO 21501-4 and JIS B9921. They are
subject to change without notice.
Specification
Description
Size Range
........................................ 0.3 to 25 µm
Channel Sizes
...................................
Standard:
0.3, 0.5, 0.7, 1.0, 2.0, 5.0 μm
Standard:
0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm
Standard:
0.3 to 10 μm, user-selectable; factory-calibrated at 0.3, 0.5, 1.0, 3.0,
5.0, 10.0
μm. Additional channel sizes available
Size Resolution
................................
<15% at
0.5 μm (per ISO 21501-4 requirements)
Counting Efficiency
.......................... 50% at
0.3 μm; 100% for particles > 0.45 μm (per JIS and ISO 21501-4)
Concentration Limits
....................... 5,950,000 particles/ft
3
at 10% coincidence loss
Light Source
..................................... Long life laser diode
Zero Count Level
.............................. <1 count/5 minutes (per JIS B9921 and ISO 21501-4)
Flow Rate
.......................................... 0.1 CFM (2.83 L/min) with
5% accuracy (meets JIS and ISO 21501-4
requirements)
Calibration
........................................ NIST traceable with TSI
®
calibration system
Calibration Frequency
.....................
Recommended minimum once per year
Sample Probe/Tubing
...................... Isokinetic sampling probe
Sampling Modes
............................... Manual, automatic, beep, cumulative/differential count or concentration
Sampling Time
.................................. 1 second to 99 hours
Sampling Frequency
........................ 1 to 9999 cycles or continuous
Exhaust
............................................. Internally filtered
Vacuum Source
................................ Internal pump
Communication Mode
...................... Modbus
®
TCP over Ethernet or USB
Data Storage
..................................... 10,000 sample records: includes date, time, six particle channels, flow, ID, and
sample volume; transferable via USB data download or TrakPro™ Lite software
Data Security
.................................... Password protected
Alarm/Status
..................................... Audible alarm sounds for counts over threshold values set in recipe, low battery,
and for instrument or flow errors.
Environmental Sensors
................... Optional temperature/RH probe supported
Display
.............................................. QVGA 3.5-inch (8.9-cm) touch screen with Windows
®
CE operating system
Languages
........................................ English, Spanish, German, French, Italian, Japanese, and Chinese (simplified)
Reports
.............................................. Provides Pass/Fail on FS-209E, ISO 14644-1 and EU GMP
Printer
................................................ Optional external printer supported
Summary of Contents for AEROTRAK 9306
Page 22: ...2 6 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...
Page 66: ...3 44 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...
Page 70: ...5 2 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...
Page 76: ...7 4 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...
Page 80: ...A 4 AeroTrak Handheld Airborne Particle Counter This page intentionally left blank ...
Page 81: ......