Troubleshooting
6-3
Symptom
Possible Cause
Corrective Action
BATTERY ERROR
Low battery (<10%).
Recharge battery or
connect AC cord.
Single battery in 100
L/min Model 9500, or
two batteries where one
has been only partially
charged
The 100 L/min Model
9500 requires two
batteries (within 25%
charge of each other) to
take a sample. If only 1
battery is installed, then
the unit will not start a
measurement and will
display the battery
warning. To correct,
place a second charged
battery in the instrument.
FLOW ERROR
Instrument was unable
to control flow rate [i.e.,
if sample tubing used is
too long (greater than
12 m on 28.3 and
50 L/min instruments,
8 m on 100 L/min
instrument)].
Restart measurement.
Pressure drop across
inlet may be too large.
Lower pressure drop
across inlet by using
larger diameter tubing,
less tubing, and/or adding
a bleed valve.
Inlet tubing kinked.
Straighten tubing.
Inlet not at ambient
pressure.
Do
not
subject the unit to
other than ambient
pressure conditions.
LASER POWER /
DETECTOR WARNING
Excessive direct light is
entering the aerosol
inlet.
Remove instrument from
direct light.
Optical path blocked.
Return to factory for
service.
Nozzle is misaligned.
Fiber attached on the
nozzle tip.
Contact TSI and return to
factory.
Detector board
damaged. Laser power
is normal.
Return to factory or
factory authorized service
centers for service.
Summary of Contents for 9510
Page 2: ......
Page 6: ......
Page 18: ...1 6 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...
Page 62: ...3 30 AEROTRAK Portable Airborne Particle Counter The standard reports are shown below ...
Page 72: ...5 4 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...
Page 76: ...6 4 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...