![TSI Instruments 9510 Operation Manuals Download Page 58](http://html1.mh-extra.com/html/tsi-instruments/9510/9510_operation-manuals_1178114058.webp)
3-26
A
ERO
T
RAK
™ Portable Airborne Particle Counter
Data Tab
The Data tab lets you preview data that has been collected. Use
the scroll bar (slide) on the right to scroll though the records. The
record number is displayed at the bottom of the tab. As each
record displays, its data and relevant parameters are displayed.
Field
Description
#, ft
3
, m
3
Button used to change between counts and
concentration displays.
Size
Channel size.
∆
Differential concentration.
Σ
Cumulative concentration.
Export the data to a flash drive. See
Print data to the optional printer. See
below.
Location
Location where the data was collected.
Sample
Duration of the sampling period.
Date
Date on which the data was collected.
Time
Time at which data was collected.
Temperature
Temperature at the end of the time the data was
collected (if probe connected during sampling).
Humidity
Humidity level at the end of the time the data was
collected (if probe connected during sampling).
Flow
Status of the flow. Options are: OK or ALRM. OK
indicates the flow rate is good; ALRM indicates flow
rate is below the defined setting.
Alarm
Alarm threshold was triggered (YES) or not
(NONE).
Laser
Status of the laser. Either: OK or SRVC.
Vol
Volume of air that was sampled.
Summary of Contents for 9510
Page 2: ......
Page 6: ......
Page 18: ...1 6 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...
Page 62: ...3 30 AEROTRAK Portable Airborne Particle Counter The standard reports are shown below ...
Page 72: ...5 4 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...
Page 76: ...6 4 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...