background image

 

 
 
 
 

QUICK START GUIDE 

ELECTROSTATIC CLASSIFIER 

MODEL 3082 

SMPS

 SPECTROMETER 

MODEL 3938 

 

Use this guide to quickly set up your Model 3082 
Electrostatic Classifier. For more detailed information, 
see the Operation and Service manual.  

Safety 

Advisory and identification labels (described in the 
Operation and Service manual) are attached to the 
exterior of the Classifier.  

S A F E T Y   W A R N I N G S

 

  High-voltage is accessible within this instrument. 

Unplug the power source before removing the cover. 

  The Model 3077 and 3077A Neutralizers contain 

radioactive material that is subject to the regulations 
of the U.S. Nuclear Regulatory Commission as well 
as local regulations.  

  The Model 3088 Advanced Aerosol Neutralizer 

contains an x-ray source. During normal operation, 
you will not be exposed to x-ray radiation. 

Do

 

not

 

make any changes (such as detaching the x-ray 
source from the neutralizer tube) to the neutralizer.  
Where required, obey local regulations to operate an 
x-ray device before turning the Model 3088 on. 

  The classifier can weigh up to 58 lb (26 kg) with the 

DMA installed. Protect your back when lifting. 

Do not

 grip the DMA when lifting. 

  A Condensation Particle Counter (CPC) may be 

combined with the classifier as particle detector. 
CPCs are Class 1 laser-based instruments. You will 
not be exposed to laser radiation during normal 
operation of the CPC. However, you must 

not

 

remove any parts from the CPC unless the manual 
instructs you to do so. 

Do not 

remove any part of 

the instrument housing while the instrument is 
powered on. 

Unpacking 

Carefully unpack the classifier. Verify that all items in 
the packing list (see below) are present. Contact TSI if 
items are missing or broken. Complete details of the 

3082 Accessory Kit, DMA Upgrade Kits, and optional 
accessories are given in the Operation and Service 
Manual.  

Retain the packaging for use when shipping the 
instrument; use of other packaging for shipment may 
result in damage to the instrument.  

Packing List 

Qty 

Description 

Model 3082 Electrostatic Classifier 

Model 3082 & Model 3938 Operation and 
Service Manual 

Model 3082 & Model 3938 Quick Start Guide 

3082 Accessory Kit 

Model 3082 Certificate of Calibration 

Cable, AC power 

Setting Up the Electrostatic Classifier 

For ventilation, install the cabinet with > 2-inch (50 mm) 
clearance between the back and side panels and any 
other surface. Setting up the classifier requires the 
following procedures (described in detail in the Operation 
and Service Manual P/N 6006760):  

1.  If needed, install the optional lead shielding. 

2.  Install a Model 3077, 3077A, or 3088 Neutralizer. 

3.  If needed, remove inlet adapter and install impactor. 

4.  Upgrade DMA if needed. 

5.  Install a DMA. Connect the DMA flow ports and high 

voltage cable to the classifier. 

6.  If needed, install the flow equalizer assembly as 

described in the Operation and Service Manual. 

7.  Connect the classifier to a PC (if used with Aerosol 

Instrument Manager

®

 software). 

8.  If setting up the classifier in an SMPS

™ 

spectrometer configuration, connect the classifier to 
a CPC.  

9.  Connect the AC power cable.  

Summary of Contents for 3082

Page 1: ...structs you to do so Do not remove any part of the instrument housing while the instrument is powered on Unpacking Carefully unpack the classifier Verify that all items in the packing list see below are present Contact TSI if items are missing or broken Complete details of the 3082 Accessory Kit DMA Upgrade Kits and optional accessories are given in the Operation and Service Manual Retain the pack...

Page 2: ...ved in Step 1 Installing Model 3088 Neutralizer The 3088 Advanced Aerosol Neutralizer has the neutralization capacity of the 3077A neutralizer but has an x ray source instead of a Kr 85 source which can be turned on and off To install 1 Remove the flag assembly 2 screws from the top of the classifier and store in Accessory kit 2 Remove the black vinyl end caps from the 3088 and install the 3088 in...

Page 3: ...igh voltage connector into the high voltage port on the classifier Align the flattened side of the connector with the flattened side of the port Note For alternative installation of the 1nm DMA to minimize diffusion losses between the Model 3777 Nano Enhancer and the 1nm DMA refer to the Operation and Service Manual Connecting Classifier to Detector The Model 3082 supports communication to a 3772 ...

Page 4: ...ench to see a description of the error condition Setting Up a Basic SMPS Scan Although the correct scan settings for your application depend on the size range you are interested in and the type of measurement you are trying to make the steps below can provide a starting point 1 Follow steps in this QuickStart guide for connecting a CPC and neutralizer to the 3082 Verify that they are listed on the...