P/N 6006761 Rev. C 

© 2016 TSI Incorporated 

TSI Model 3082 Quick Guide page 

2

 

10.  Upgrade CPC firmware if necessary. 

Installing Optional Lead Shielding 

In some countries or localities, the Model 3077A 
neutralizer must be operated with lead shielding 
installed. It is an optional accessory. To install:

 

1.  Remove the instrument cover.

 

2.  Remove the flag assembly and the flag assembly 

housing.

 

 

3.  Locate the four lead shielding guide rods and slide 

the lead shield into place around them. 

 

 

4.  Replace the flag assembly housing, flag assembly, 

and cover.

 

Installing Model 3077/3077A Neutralizer 

The neutralizer provides a known charge distribution 
on the aerosol entering the DMA. Model 3077 (2 mCi) 
and Model 3077A (10 mCi) have a radioactive Krypton 
85 source. To install: 

1.  Remove the flag assembly (2 screws) from the top 

of the classifier. 

2.  Fit the long tube on the Model 3077/3077A 

Neutralizer into the flag assembly.  

3.  Remove the 6006269 retainer bracket from 

accessory kit.  

4.  Slide the bracket over the bottom tube. Compress 

the spring and push the bracket into place. 

 

5.  Install the flag assembly into the housing and replace 

the two screws removed in Step 1. 

Installing Model 3088 Neutralizer 

The 3088 Advanced Aerosol Neutralizer has the 
neutralization capacity of the 3077A neutralizer but has 
an x-ray source (instead of a Kr-85 source) which can be 
turned on and off. To install:  

1.  Remove the flag assembly (2 screws) from the top of 

the classifier and store in Accessory kit. 

2.  Remove the black vinyl end caps from the 3088 and 

install the 3088 in the Electrostatic Classifier.  

 

3.  Replace the two screws removed in Step 1. 

Installing Impactor 

To remove particles larger than the size range of interest 
and/or to measure aerosol inlet flow to the classifier, the 
inlet adapter should be removed and an impactor should 
be installed. The impactors are provided in the 3082 
Accessory Kit (shipped with the instrument). To install: 

1.  Remove impactor and impaction plate from 

accessory kit. Push the stainless steel impactor plate 
into the body. 

 

2.  Remove the inlet adapter and put it in the Impactor 

Kit case. 

3.  Push the impactor into position. Lock in place by 

twisting clockwise until the dot on the impactor aligns 
with the arrow on the bezel.  

Upgrading DMA 

If you have a previous-model DMA (3081 or 3085) 
purchased before 2014, you can purchase an upgrade 
kit (3081U, 3085U) from TSI to make it compatible with 
the Model 3082. See the 3082 manual for details on how 
to perform the upgrade. 

Summary of Contents for 3082

Page 1: ...structs you to do so Do not remove any part of the instrument housing while the instrument is powered on Unpacking Carefully unpack the classifier Verify that all items in the packing list see below are present Contact TSI if items are missing or broken Complete details of the 3082 Accessory Kit DMA Upgrade Kits and optional accessories are given in the Operation and Service Manual Retain the pack...

Page 2: ...ved in Step 1 Installing Model 3088 Neutralizer The 3088 Advanced Aerosol Neutralizer has the neutralization capacity of the 3077A neutralizer but has an x ray source instead of a Kr 85 source which can be turned on and off To install 1 Remove the flag assembly 2 screws from the top of the classifier and store in Accessory kit 2 Remove the black vinyl end caps from the 3088 and install the 3088 in...

Page 3: ...igh voltage connector into the high voltage port on the classifier Align the flattened side of the connector with the flattened side of the port Note For alternative installation of the 1nm DMA to minimize diffusion losses between the Model 3777 Nano Enhancer and the 1nm DMA refer to the Operation and Service Manual Connecting Classifier to Detector The Model 3082 supports communication to a 3772 ...

Page 4: ...ench to see a description of the error condition Setting Up a Basic SMPS Scan Although the correct scan settings for your application depend on the size range you are interested in and the type of measurement you are trying to make the steps below can provide a starting point 1 Follow steps in this QuickStart guide for connecting a CPC and neutralizer to the 3082 Verify that they are listed on the...

Reviews: