Chapter 5
Troubleshooting
Table 5 lists the symptoms, possible causes, and recommended solutions for
common problems encountered with the V
ELOCI
C
ALC
. If your symptom is
not listed, or if none of the solutions solves your problem, please contact
TSI.
Table 5. Troubleshooting the V
ELOCI
C
ALC
Symptom
Possible Causes
Corrective Action
No display
Unit not switched on
Switch on the unit.
Low or dead batteries
Replace the batteries or
plug in the AC adapter.
Dirty battery contacts
Clean the battery
contacts.
BAT is blinking
Batteries are low
Replace or recharge
batteries.
Display reads
"LO BAT"
Low battery charge
Replace or recharge
batteries.
Wrong AC adapter
Replace with the correct
AC adapter.
Low AC line voltage
Correct the AC line
voltage or use batteries.
Dirty battery contacts
Clean the battery
contacts.
Temperature
initially reads
high
Temperature sensor is
still warm from velocity
mode
Allow about 30 seconds
before reading
temperature.
Display reads
"ERR"
You are trying to enter
more than 255 readings
Read or record the
average; clear the
storage register and
proceed.
Display reads
"CAL"
The V
ELOCI
C
ALC
has
detected an internal fault
Return to factory for
service.
Velocity reading
fluctuates badly
The flow is fluctuating
Reposition the probe in a
less turbulent section of
the flow or use a longer
time constant.
Display says
"OVER"
The velocity,
temperature or pressure
is too high
Use an alternate
measurement method.