Appendix E
Timing in Fit Test Mode
Test Sequence
When performing a fit test, PortaCount
®
Pro
Respirator Fit Tester
automatically samples both the ambient air and the particles inside
the respirator. An internal switching valve automatically switches
between the two sampling ports. In each test cycle, (exercise) both an
ambient sample and a respirator sample are taken. Also, time is
allowed for purging of the sensor between each of the samples.
The PortaCount Pro
fit tester uses the ambient sample taken before
and after each exercise to compute the fit factor. The two ambient
particle concentrations are averaged together and then divided by the
mask particle concentration to yield the fit factor for that exercise.
Averaging the before and after ambient particle concentrations is
done to offset any ambient particle concentration drift that may have
occurred during the mask sample.
C a u t i o n
Do
not
lengthen the Twin Tube Assembly more than a
few inches unless the mask sample purge time has also
been extended accordingly. Insufficient purging between
mask and ambient samples will result in unrealistically
low fit factors. Shortening the Twin Tube Assembly is
never a problem.
Summary of Contents for PORTACOUNT PRO+ 8038
Page 2: ......
Page 10: ...4 Introduction and Overview This page intentionally left blank ...
Page 22: ...16 Chapter 2 This page intentionally left blank ...
Page 30: ...24 Chapter 3 This page intentionally left blank ...
Page 82: ...76 Chapter 7 This page intentionally left blank ...
Page 86: ...80 Appendix A This page intentionally left blank ...
Page 90: ...84 Appendix B This page intentionally left blank ...
Page 94: ...88 Appendix D This page intentionally left blank ...
Page 96: ...90 Appendix E This page intentionally left blank ...
Page 100: ...94 Appendix F This page intentionally left blank ...
Page 101: ...95 Appendix G The Rainbow Passage ...
Page 102: ...96 Appendix G This page intentionally left blank ...
Page 106: ......