Troubleshooting
75
Symptom
Possible Causes
Solution
High
Concentration
message
A near-by particle
generator is elevating
ambient concentration
levels.
Turn off or turn down any
particle generators
operating in the vicinity
such as humidifiers or the
Model 8026 Particle
Generator.
Keep all forms of particle
generation at least 6 ft
(1.8 m) from the
PortaCount fit tester
during fit testing.
The room ventilation
system is reduced or
turned off.
Increase room ventilation.
Ambient concentration is
naturally elevated,
In situations where the
ambient concentration is
elevated through pollution,
construction, or other
means where it cannot be
controlled, the trigger
threshold for the warning
may be raised as
described in
Operating Modes and
User Interface
. Note that
prolonged exposure to
excessive particle
concentration levels could
degrade the performance
of the PortaCount fit tester
over time, requiring more
frequent maintenance.
Summary of Contents for PORTACOUNT PRO+ 8038
Page 2: ......
Page 10: ...4 Introduction and Overview This page intentionally left blank ...
Page 22: ...16 Chapter 2 This page intentionally left blank ...
Page 30: ...24 Chapter 3 This page intentionally left blank ...
Page 82: ...76 Chapter 7 This page intentionally left blank ...
Page 86: ...80 Appendix A This page intentionally left blank ...
Page 90: ...84 Appendix B This page intentionally left blank ...
Page 94: ...88 Appendix D This page intentionally left blank ...
Page 96: ...90 Appendix E This page intentionally left blank ...
Page 100: ...94 Appendix F This page intentionally left blank ...
Page 101: ...95 Appendix G The Rainbow Passage ...
Page 102: ...96 Appendix G This page intentionally left blank ...
Page 106: ......