X
X
L
L
D
D
C
C
M
M
A
A
N
N
.
.
d
d
o
o
c
c
3
3
-
-
5
5
2
2
R
R
E
E
V
V
E
E
Either the Start Time or the Stop Time may be replaced with a semicolon, in which case that item will
be unchanged by the command.
Sample entry:
F26 ; XXX:XX:10:00.000<CR>
Response:
OK<CR><LF>
This entry will leave the Start Time set to whatever it was before, and will change the Stop Time to
10 minutes after the hour. If both the Start Time and the Stop Time match the current time, then the
pulse is turned off.
Sample entry:
F26 XXX:XX:XX:XX.2XX XXX:XX:XX:XX.XXX<CR>
Response:
OK<CR><LF>
The pulse would normally be turned on every second at 200 milliseconds, but the Stop Time also
matches then, so the pulse will stay off.
To clear the output regardless of the time: F26 000:00:00:00.000 XXX:XX:XX:XX.XXX
To set the output regardless of the time:
F26 XXX:XX:XX:XX.XXX 000:00:00:00.000
The default power-on Pulse Out setting is: F26 000:00:00:00.00 000:00:00:00.000
Since the day is day 000, the Start and Stop times never match.
When the power is turned off, the current setting is saved in non-volatile memory, to be restored when it
is next turned on.
3-227 reserved
3-228 SERIAL FUNCTION F28 - TIME INTERVAL/EVENT TIMING INPUT (OPTION)
Refer to manual section 1-41 for specifications on the Time Interval/Event Timing Input. Use Serial
Function F28 to request a time interval measurement or event times. Send a string of the form:
F28<SEP><MODE><CR>
where:
F
=
ASCII character F
28
=
function number
<SEP>
=
one or more optional separator characters; either space, comma
or tab
<MODE>
=
either TI for Time Event or ET for Event Timing (this field is
optional with the default being ET)
<CR>
=
carriage return character
If the mode was TI, the port will respond with a string of the form:
<NSEC><CR><LF>
Summary of Contents for 151-600
Page 2: ...Model XL DC Serial Number__________________ September 15 1997 Revision E...
Page 4: ......
Page 27: ...X...