![Troxler RoadReader Plus 3450 Manual Of Operation And Instruction Download Page 101](http://html1.mh-extra.com/html/troxler/roadreader-plus-3450/roadreader-plus-3450_manual-of-operation-and-instruction_1172544101.webp)
6–
The gauge displays the thin layer special calibration name and both
B values. To return to the
Thin Lift Special
menu, press the
〈
ENTER
〉
key. Return to the
Ready
screen by pressing the
〈
ESC
〉
key.
Create a Thin Layer Special Calibration
To create a thin layer special calibration, the operator must take
readings with the gauge and enter the core density. The gauge uses
these values to calculate the thin layer special calibration constants.
The Model 3450 can store one partial calibration. A partial
calibration allows the operator to enter the asphalt parameters and
take readings at different times.
METHOD 1 – MEASURE DENSITY FIRST.
To create a thin
layer special calibration, press
〈
4
〉
New
at the
Thin Lift Special
menu
shown on page 6–16. The gauge displays the
Thin Special
Calibration
menu:
If a partial calibration was saved earlier, the gauge will display a
P
beside the option entered before (see page 6–21). To take readings
for the thin layer special calibration, press
〈
1
〉
and see
Measuring
the Density
below. To enter the core density, press
〈
2
〉
and see
Entering the True Densities
on page 6–20.
Measuring the Density.
Prepare the test site as described on
page 4–6. Place the gauge on the test site. Lower the source rod to
the backscatter position. Release the trigger and lock the source rod
into position. You should hear a
click
when the source rod locks into
position.
Thin Special Calib
1-Measure Density
2-Input True Density
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...