Appendix A–20
Table A–1. Radiation Profile for Model 3450 Gauge
(Dose Rate in Millirem per Hour)
Surface
5 cm
30 cm
100 cm
G
a
mma
N
e
ut
ron
Tot
a
l
G
a
mma
N
e
ut
ron
Tot
a
l
G
a
mma
N
e
ut
ron
Tot
a
l
G
a
mma
N
e
ut
ron
Tot
a
l
GAUGE
Front
6.5 N/A 6.5 3.5 N/A 3.5 0.8 0.3 1.1 0.1 0.1 0.2
Back
14.0 N/A 14.0 7.0 N/A 7.0 0.7 0.5 1.2 0.14 0.1 0.24
Left Side
11.0 N/A 11.0 3.5 N/A 3.5 0.8 0.1 0.9 0.1 0.0 0.1
Right Side
3.5 N/A 3.5 2.5 N/A 2.5 1.0 0.3 1.3 0.25 0.07 0.32
Top
14.0 N/A 14.0 6.0 N/A 6.0 1.7 0.35 2.05 0.3 0.1 0.4
Bottom
14.0 N/A 14.0 10.0 N/A 10.0 1.6 0.25 1.85 0.2 0.05 0.25
GAUGE IN TRANSPORT CASE
Front
7.0 N/A 7.0 3.0 N/A 3.0 0.5 0.1 0.6 0.12 0.0 0.12
Back
2.0 N/A 2.0 1.5 N/A 1.5 0.9 0.3 1.2 0.2 0.05 0.25
Left Side
0.2 N/A 0.2 0.15 N/A 0.15 0.08 0.5 0.58 0.0 0.05 0.05
Right Side
5.0 N/A 5.0 2.5 N/A 2.5 0.5 0.55 1.05 0.15 0.1 0.25
Top
5.0 N/A 5.0 4.0 N/A 4.0 1.0 0.35 1.35 0.15 0.05 0.2
Bottom
5.0 N/A 5.0 3.5 N/A 3.5 0.85 0.25 1.1 0.15 0.1 0.25
Survey Equipment Used:
Gamma:
Ludlum Model 14C Survey Meter, serial #57418,
calibrated 10/02/1996.
Neutron:
Snoopy Portable Neutron Doserate Meter, model #NP-2,
serial #NP183404, calibrated 12/07/1995.
NOTES:
1. All tabulated values are in units of mrem/hr.
2. The orientation of the gauge in the transport case is as follows:
a. Back of gauge to front of case.
b. Bottom of gauge to right side of case
c.
Top of gauge to left side of case
3. Dose rates measured by the State of North Carolina Department of
Environment and Natural Resources, Division of Radiation Protection.
4. Dose equivalent rates for 8 mCi Cs-137 and 40 mCi Am-241:Be.
5. Surface indicates surface of the gauge or surface of the transport case.
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...