Model 3450
Appendix C–15
ERROR! INCORRECT BASEBOARD ID
This error occurs if the ID stored in the base unit does not match the
ID stored in the control unit.
Solution
: Make sure the control unit is the correct unit for a
particular base unit.
INVALID DEPTH – CHECK SOURCE ROD
This error occurs when the source rod is not at a valid depth when
starting a measurement.
Solution
: Check source rod position to ensure it is at a valid depth
when starting a measurement. Recalibrate depth strip. Call the
nearest Troxler Service Center.
KEYPAD ERROR!
This error is reported after the power-up self-test if a stuck key is
detected.
Solution
: Clear the keypad of any objects pressing keys on
powerup.
LCD ERROR!
During the Liquid Crystal Display (LCD) test, a sequence of
characters is written to the LCD and is then read back. This message
occurs if a character read back does not match the character written
out.
Solution
: Call the nearest Troxler Service Center.
C. T
RO
UB
LE
SH
OO
TIN
G
Summary of Contents for RoadReader Plus 3450
Page 7: ...Model 3450 vii NOTES...
Page 13: ...xii NOTES...
Page 17: ...1 4 GAUGE PARTS AND ACCESSORIES Figure 1 1 Model 3450 Parts and Accessories...
Page 23: ...2 4 Figure 2 1 Effect of Moisture on Depth of Measurement...
Page 53: ...3 28 NOTES...
Page 169: ...Appendix E 2 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 171: ...Appendix E 4 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 173: ...Appendix E 6 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 175: ...Appendix E 8 STANDARD COUNT LOG Gauge Serial Number Date Moisture MS Density DS1 Density DS2...
Page 181: ...Appendix G 4 NOTES...
Page 187: ...Appendix H 6 NOTES...
Page 199: ...Index 12 NOTES...