3–8
The gauge displays:
<STORE> for Next
<ENTER> to Calculate
Press the
〈
STORE
〉
key to store the values and to enter those for the
next location. When the measured and actual densities for each
location have been entered, press the
〈
ENTER/START
〉
key to
calculate the mix calibration. The gauge displays:
Slope = #.##
Intercept = ###.#
Do you want to
accept this mix?
To return to the
Setup
menu without saving the mix calibration,
press the
〈
NO
〉
key. To save the mix calibration, press the
〈
YES
〉
key. The gauge requests a mix calibration ID as shown:
Mix Calib. ID
<ALPHA> for Letters
Press <ENTER>
Press the
〈
ALPHA LOCK
〉
key to enable the alphabetic keys on the
gauge. When the alphabetic keys are enabled, the symbol
A
appears
in the upper right of the display, as shown above.
Enter a unique mix calibration ID and press
〈
ENTER/START
〉
. The
gauge displays:
Select Memory Cell
1. Empty 2. Empty
3. Empty 4. Empty
Press # To Select