Trotec BA16WP Operating Manual Download Page 7

EN

7

vane anemometer controlled via smartphone BA16WP

Continuous measurement

Info

Note that moving from a cold area to a warm area can
lead to condensation forming on the device's circuit
board. This physical and unavoidable effect can falsify
the measurement. In this case, the app will either
display incorrect measured values or none at all. Wait a
few minutes until the device has become adjusted to
the changed conditions before carrying out a
measurement.

When the appSensor has been successfully connected to the
terminal device, a continuous measurement is started and
indicated. The refresh rate is 1 second. The 12 most recently
measured values are displayed graphically (9) in chronological
sequence. The currently determined and calculated measured
values are displayed numerically (10).

Sensors

MEASUREMEN

TS

Reports

CUSTOMERS

SETTINGS

Temperatur

e

Relative humidity

[24,7 | 25,4]    Ø 25

[37,6 | 58]    Ø 40

25,3 °C

37,6 % RH

°C

% RH

% RH

°C

9

10

25

26

27

28

40

35

11

Air velocity

0,43 m/s

Volumetric flow

0,0345 m³/min

BA16WP

m/s

[0,42 | 0,44]    Ø 0,43

No. Designation

Meaning

9

Relative humidity

indication as
graphic display

Indicates the relative humidity as
measured over the course of time.

Temperature

indication as
graphic display

Indicates the temperature as
measured over the course of time.

Volumetric flow

indication as
graphic display

Indicates the volumetric flow as
calculated over the course of time
(after having entered the area).

No. Designation

Meaning

10 Tabular 

value

display

Temperature:

 Indicates minimum,

maximum and average value as well
as the current value.

Relative humidity:

 Indicates

minimum, maximum and average
value as well as the current value.

Air velocity:

 Indicates minimum,

maximum and average value as well
as the current value.

Enter area / volumetric flow:

 Tap on

the field an enter the cross-sectional
area. Then the calculated value for the
volumetric flow will be displayed.

11

Menu

 button

Opens the menu to adjust the settings
of the current measurement.

Info

The indicated measured values will not be saved
automatically.

Info

By tapping on the graphic display (9) you can switch
over to the numeric display and vice versa.

Info

The volumetric flow can only be approximated, since
the flow type and speed of the air current is not
homogenous and thus identical across the entire
cross-section of the line / pipe but varies from the
centre to the periphery! For a preferably precise
measurement move the sensor slowly and evenly in a
spiral motion from the centre to the edge and use the
determined average value as result.

Summary of Contents for BA16WP

Page 1: ...TRT BA BA16WP TC2019 16 001 EN BA16WP EN OPERATING MANUAL VANE ANEMOMETER CONTROLLED VIA SMARTPHONE...

Page 2: ...he device Always store the manual in the immediate vicinity of the device or its site of use Warning Read all safety warnings and all instructions Failure to follow the warnings and instructions may r...

Page 3: ...ve a minimum distance of 20 cm between the pacemaker and the device Also do not use the device near automatically controlled systems such as alarm systems and automatic doors Radio waves may interfere...

Page 4: ...n use report templates and ready made text blocks for various fields of application to turn the data into professional reports Device depiction BA16WP 1 2 3 4 5 No Designation 1 Measuring sensor 2 LED...

Page 5: ...protected from frost and heat protected from dust and direct sunlight The storage temperature is the same as the range given in the Technical data chapter Remove the batteries from the device Operati...

Page 6: ...e appSensor to the terminal device The Trotec MultiMeasure Mobile app is installed The Bluetooth function on your terminal device is activated 1 Start the Trotec MultiMeasure Mobile app on the termina...

Page 7: ...of time Temperature indication as graphic display Indicates the temperature as measured over the course of time Volumetric flow indication as graphic display Indicates the volumetric flow as calculate...

Page 8: ...measurement button 13 to switch from continuous measurement to individual value measurement Individual value measurement has been selected as measuring mode Return to the screen displaying the measur...

Page 9: ...Relative humidity 24 7 25 4 25 37 6 58 40 37 6 RH 25 3 C Stop recording m s m s BA16WP No Designation Meaning 20 REC button Opens the sensor settings menu 21 Stop recording button Stops the current re...

Page 10: ...nsors MEASUREMEN TS Reports CUSTOMERS SETTINGS test_a1 Basic data Evaluations Evaluation parameters Values Generate table Generate graphic 29 30 31 32 33 34 No Designation Meaning 29 Basic data button...

Page 11: ...Press the corresponding button 37 to display the desired report An input mask opens in which you can view and edit all the information Sensors MEASUREMEN TS Reports CUSTOMERS SETTINGS Reports 03 26 20...

Page 12: ...pair Do not modify the device or install any spare parts For repairs or device testing contact the manufacturer Errors and faults The device has been checked for proper functioning several times durin...

Page 13: ...our website www trotec24 com The separate collection of waste electrical and electronic equipment aims to enable the re use recycling and other forms of recovery of waste equipment as well as to prev...

Page 14: ...Trotec GmbH Grebbener Str 7 D 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com...

Reviews: