INSTALLATION & OPERATING DATA
10 APPROVALS AND CERTIFICATION
10.2
Europe (ATEX)
TX6373.01.xx Toxic Gas Sensor / Transmitter (Group I)
Ex Certificate number:
SIRA 02ATEX2052X
Ex Certification code:
I M1 Ex ia I Ma (Ta = -20
°
C to +60
°
C)
TX6373.02.12 Toxic Gas Sensor / Transmitter (Group II)
Ex Certificate number:
SIRA 02ATEX2052X
Ex Certification code:
II 1G Ex ia IIC T4 Ga (Ta = -20
°
C to +60
°
C)
Special Conditions for Safe Use
When the TX6373 is specifically required for group II, category 1 equipment, the
TX6373 should not be installed in a location where the external conditions are
conducive to the build-up of electrostatic charge as the polycarbonate window may
store an ignition-capable level of electrostatic charge. Additionally, the equipment
should only be cleaned with a damp cloth.
General Conditions for Safe Use
Prior to installation, it is essential that user refers to the above certificate to ensure
that the termination and cable parameters are fully complied with and are
compatible with the application. Copies of certificates are available from Trolex.
ATEX Directive (94/9/EC)
Australia/New Zealand (ANZEx)
TX6373-series Toxic Gas Sensor / Transmitter
Ex Certificate Number:
ANZEx 12.3016X
Ex Certification Code:
Ex ia I (Ta = -20
°
C to +60
°
C)
Ex ia IIC T4 (Ta = -20
°
C to +60
°
C)
Special Conditions for Safe Use:
Prior to installation, it is essential that the user refers to the above certificate to
ensure that the termination and cable parameters are fully complied with and are
compatible with the application. Copies of certificates are available from Trolex.
When the TX6373 is specifically required for group II, Zone 0, the TX6373 should
not be installed in a location where the external conditions are conducive to the
build-up of electrostatic charge as the polycarbonate window may store an ignition-
capable level of electrostatic charge. Additionally, the equipment should only be
cleaned with a damp cloth.
10.1
20/25
ISSUE N 06/15