INSTALLATION & OPERATING DATA
5 INSTALLATION
continued
5.8 Operating Life
Electrochemical cells contain an electrolyte that is gradually consumed during
use. The average life is about two years, dependent upon the duty cycle.
The response should be checked at regular intervals.
5.9 Sensitivity
Electrochemical cells for toxic gases can be affected by other interfering
gases which may displace the subject gas being monitored. Steam laden
atmospheres and condensation can also reduce the sensitivity.
5.10 Flammable
Be aware that some toxic gases are also ‘flammable’ at high percentage
concentrations.
5.11 Biased Sensors
Some gas sensors must be continuously powered to maintain the calibration.
If the gas head is removed from any supply voltage for greater than 10 minutes,
it could take 24–48 hours to restore its calibration.