TX6351 • TX6352 User Manual
34 P5536.1600 rev B
P5536.1600 rev B 35
Main Menu
Press the R key to enter the Main Menu
Checkpoint
The front cover of Sentro may be safely
removed, for setup in a hazardous area, even
with the power applied.
4.3 Security Code Access
A security code may be requested:
Press the
R
key to change the digit
Use the
L
key to increment the digits
Use the
R
key to select Enter and return to
the Main Menu
Section 4.4
Section 4.3
Section 4.4
4.4 Main Menu
Prompt Bar
Calibrate
Sentro 1 Setup
Output Setup
Modbus Setup
Module Setup
Checkpoint
Sentro will automatically return to
the Main Display if no commands are
given within 1 minute. The time limit is
extended to 8 minutes during calibration
to allow the gas value to stabilise.
Section 4.4.1
Section 4.4.2
Section 4.4.3
Section 4.4.4
Section 4.4.5
Zero
Set Test Gas
Span
System Information
Display Setup
Alert Setup
Set Security Code
Exit
Output Information
Trim Output Zero
Trim Output Span
Reset Output
Set Relay 1 Mode
Set Relay 2 Mode
Exit
Modbus Address
Baud Rate
TX On Delay
TX Off Delay
Exit
Setpoint 1
Setpoint 2
STEL/TWA
Exit
(L)
(R)