TX6351 • TX6352 User Manual
38 P5536.1600 rev B
P5536.1600 rev B 39
Calibrate Span using a Test Gas of DIFFERENT value
The display will show the concentration
of Span Gas that is EXPECTED to be
used.
Set the expected value to agree with
the actual Test Gas value:
•
Follow
the
prompt
bar
and screen instructions
Press
R to continue with the
normal span calibration
(previous page)
4.4.2 Sentro 1 Setup
The basic functional characteristics of the
sensor can be setup, or reviewed, from a
choice of parameters.
System Information Display
Display Setup
• The display backlight illumination
may be set to on, or off , to reduce
power consumption.
(TX6351 only)
• The contrast of the display
may
be
set
for
best
visual appearance.
www.trolex.com
www.trolex.com