4
2.
Specifications
Safety Classification
Electric
protection
Type
Class I
Electric
protection
Level
Defibrillation proof type CF applied Part
Ingress Protection
IP34 (Prevent solid objects larger than 2.5mm in diameter and water intrusion
from splashing in all directions)
Working mode
Continuous
Classification
Portable equipment, non-portable syringe pump
Specification Parameters
Compatible Syringes 3ml (most major brands), 5/6ml, 10/12ml, 20ml, 30/35ml, 60ml
System Accuracy
≥
1ml/h, ±3%
<
1ml/h, ±5%
Infusion Rate
Syringe size 3ml: (0.1-100) ml/h
Syringe size 5/6ml: (0.1-150) ml/h
Syringe size 10/12ml: (0.1-300) ml/h
Syringe size 20ml: (0.1-600) ml/h
Syringe size 30/35ml: (0.1-900) ml/h
Syringe size 60ml: (0.1-1500)ml/h
Bolus Rate
Syringe size 3ml: (0.1-100) ml/h
Syringe size 5/6ml: (0.1-150) ml/h
Syringe size 10/12ml: (0.1-300) ml/h
Syringe size 20ml: (0.1-600) ml/h
Syringe size 30/35ml: (0.1-900) ml/h
Syringe size 60ml: (0.1-1500)ml/h
Bolus preset value
Min: 0.1ml Max: max rate of accordingly loaded syringe size
KVO Rate
0-5.00ml/h
Micro mode
setting range
Syringe size 3ml: (100-100) ml/h
Syringe size 5/6ml: (100-150) ml/h
Syringe size 10/12ml:(100- 300) ml/h
Syringe size 20ml: (100-600) ml/h
Syringe size 30/35ml: (100-900) ml/h
Syringe size 60ml: (100-1500) ml/h
Minimum flow rate
increment
0.01ml/h
VTBI
0-9999.99ml, minimum step is 0.01ml
Total
Volume
Infused
0-9999.99ml, minimum step is 0.01ml