Finish Pre-alarm
Time for pre-alarm refers to the time of activating near completion alarm when the fluid/drug
infused volume is nearly reaching the preset value.
Click
『
Finish pre-alarm
』
to enter into the time for pre-alarm setting interface. Click the preset
time option to set the finish pre-alarm time.
Micro Mode
Click
『
Micro mode
』
to enter into micro mode setting interface. ON/OFF is optional in this
function. Under the ON mode, set the rate limit, then the infusion rate under any infusion mode is
not allowed to exceed this limit.
Syringe Size
Max Rate Range
3ml
100-100ml/h
5/6ml
100-150ml/h
10/12ml
100-300 ml/h
20ml
100-600 ml/h
30/35ml
100-900 ml/h
60ml
100-1500 ml/h
Reset Total Volume
Click
『
Reset total volume
』
, the interface displays the operation confirming prompt box, click
『
Yes
』
to confirm reset, otherwise, please click
『
No
』
.
General
In the main interface, click
『
General
』
to enter into the General equipment setting interface.
Date & Time
Click
『
Date &Time
』
to enter into the date and time setting interface. It allows to set the date, time
and format in this interface.
When setting date and time, directly input the numerical value in the input method interface. For
example, to preset date “10-24-2019”, input “10242019”; to preset the time “13: 34”, input
“1334”.
The time can be displayed in 24h format or 12h format. The date can be displayed in British type,
American type or Chinese type. Please set according to the requirement.