Warning:
When adopting fluid/drug of a high viscosity and the occlusion pressure is set at low level, it
is possible that the system will report occlusion alarm even when the line is not obstructed.
Under this condition, please carefully observe the pressure indication icon in the display
screen and infusion line, and raise the occlusion pressure if needed.
When the occlusion pressure is set to high grade, the large amount of pressure inside the
pipeline is likely to pop out the extension line connected to the syringe. Please confirm that
the extension line is securely attached to the syringe.
When the occlusion pressure is set at high level, it may possibly cause the patient discomfort.
After raising the occlusion pressure, please carefully observe the condition of the patient,
and immediately take measure if there’s any abnormality.
Under the equipment fault state, the max pressure generated by the infusion line is 900
mmHg. Under single fault state, the max infusion volume is 2ml.
Occlusion Pressure Level: 3 levels
Level
Pressure Intensity
(mmHg)
Level
Pressure Intensity
(mmHg)
Level
Pressure Intensity
(mmHg)
1
300
2
600
3
900
Pressure Unit
Click
『
Pressure unit
』
to enter into pressure unit select setting interface, four units are available:
mmHg, kPa, bar, PSI, click the preset unit option.
Note:
●
Please carefully confirm when changing the current pressure unit.
Unit Mark
Unit Conversion
kPa
1 kPa=7.5mmHg=0.145psi=0.01bar
PSI
1psi=51.713mmHg=6.895kpa=0.069bar
Bar
1bar=787.5mmHg=15.225psi=105kPa
Reminder Alarm
Click
『
Reminder alarm
』
to enter into the time for reminder alarm setting interface. Click the
preset time option to set the reminder alarm time.