Last Therapies
Click
『
Records
』
then
『
Last therapies
』
to enter last therapies interface. System will store 20 of
the previous most recent therapies to select.
8.
Alarm Prompt and Troubleshooting
Introduction to Alarm Levels
During infusion preparation and infusion, this equipment will alarm when reaching or exceeding
the set alarm threshold value and prompt with sound, light and text. According to the importance
of alarm information as well as the emergency and safety, the alarm is divided into three levels:
high, middle and low. Please refer to table below for details:
Alarm Level
Sound Signal Interval
Light color /flash frequency
High alarm
10s
Red indicator flashes /2.0±0.6Hz
Middle alarm
15s
Yellow indicator flashes / 0.6±0.2Hz
Low alarm
Once, not repeated
Yellow indicator lights on
If there’s an alarm, the system will display the alarm interface. If the alarm level is high, click
『
OK
』
, stop the alarm, and exit the alarm interface. If the alarm level is middle or low, click
『
OK
』
, the sound signal will stop, and exit the alarm interface.
Click
『
Mute
』
to mute, if alarm is not eliminated, the alarm sound will be sent out 2 minutes later.
Warning:
●
Some alarm threshold values of this equipment can be set by the user, for
example: occlusion pressure, reminder alarm, VTBI infused pre-alarm, alarm
sound volume and so on. The user shall confirm the parameters when setting
the alarm threshold value. Otherwise, it may possibly influence the alarm
function or infusion safety.
Multilevel Alarm Rules
When there’re several alarms, the system will alarm according to the following rules:
Multilevel Alarm
Rules
Several alarms of different
levels generate simultaneously
Display the alarms of highest level with sound, light and text,
the reports middle alarm after eliminating all alarms of
highest level
Several alarms of same level
generate simultaneously
Alarm circularly displays by turns, the time interval is 1s
Note: When alarming, the corresponding alarm information will display on the title of the
screen.