User’s Guide | 1200g

                                

 

 

 3 

Set Up   

PRECAUTIONS AND REQUIREMENTS 
FOR PRECISION WEIGHING  

Precision scales should be placed in a location that 
will not degrade the accuracy. 

Avoid extremes of temperature. Do not place in 
direct sunlight or near air conditioning vents. 

Avoid unsuitable tables. The tables or floor must be 
rigid and not vibrate. Do not place near vibrating 
machinery. 

Avoid unstable power sources. Do not use near 
large users of electricity such as welding equipment 
or large motors. 

Avoid high humidity that might cause condensation. 
Avoid direct contact with water. Do not spray or 
immerse the scales in water. 

Avoid air movement such as from fans or opening 
doors. Do not place near open windows. Keep the 
scales clean. 

Do not stack material on the scales when they are 
not in use. 

THE CZ PRECISION SCALE 

Level the scale by adjusting the four feet. The scale 
should be adjusted such that the bubble is centered 
and the scale does not rock from corner to corner. 

Attach the adapter to connector on the side of the 
scale, plug into standard 110VAC outlet.

 

Summary of Contents for 1200g High Precision Scale

Page 1: ...UG Version 1 2_12 30 13 1200g High Precision Scale User s Guide 1200g High Precision Scale Legal for Trade Includes RS 232 interface port...

Page 2: ...ross Road Olive Branch MS 38654 800 238 0152 www trinerscale com 2013 Triner Scale All rights reserved CALIBRATION WARNING Calibration AND inspection of calibration properties is prohibited unless don...

Page 3: ...10 Output of the Weight 11 Output to a Receipt Printer 11 Output to a PC 12 Output to a Remote Display 13 Error Codes 14 Weighing Unit Conversions 15 NOTICE The CZ 1200 High Precision scale is a very...

Page 4: ...ting features include automatic zero tracking and an accumulation facility that allows the count to be stored and recalled as an accumulated total Specifications Capacity 1200g x 0 1g NTEP Certified L...

Page 5: ...rs of electricity such as welding equipment or large motors Avoid high humidity that might cause condensation Avoid direct contact with water Do not spray or immerse the scales in water Avoid air move...

Page 6: ...erface ESC Return to normal operation when the scale is in a parameter setting mode avoids saving recently entered changes UNIT MODE UNIT Selects either grams pounds carat ounce etc for the weighing u...

Page 7: ...account for minor drifting or accumulation of material on the platform When the weight goes over 20 of the full capacity automatic zeroing is disabled TARING If necessary zero the scale by pressing th...

Page 8: ...form 2 After the stable icon displays press the PRINT key 3 ACC 1 will display 4 Remove all weight from scale platform then place next item to be weighed on the platform 5 Per 2 above press the PRINT...

Page 9: ...isplay Press the MODE key to move through the available parameters To access the settings in the parameter showing in the display such as F1 Unt press the ENTER key The first setting will display such...

Page 10: ...dwt F2 bl BACKLIGHT SETTING EL on EL backlight always on EL Au EL backlight automatically turn on EL off El backlight always off F3 cal CAUTION Calibration is to be performed by a qualified scale tech...

Page 11: ...n is to be performed by a qualified scale technician ONLY LINEAR CALIBRATION Remove all weight from the pan Press ENTER PIN will display Enter the password by pressing the keys PRINT S P Load 0 will d...

Page 12: ...can select from 600 1200 2400 4800 9600bps SELECT PRINTER OUTPUT MODE If you select P prt or P auto for printer after setting baud rate the display will show tp or LP 50 This is printer mode you can...

Page 13: ...onnect the scale to the printer using a standard DB9 null modem cable 2 Access the Technical Parameters of the CZ 1200 scale see the previous section Adjusting the Parameters 3 Navigate to the P2 Con...

Page 14: ...he CZ 1200 scale see the previous section Adjusting the Parameters 3 Navigate to the P2 Con parameter 4 Press ENTER Press MODE to select S 232 or S USb as required based on the connection 5 Press ENTE...

Page 15: ...dem cable 2 Access the Technical Parameters of the CZ 1200 scale see the previous section Adjusting the Parameters 3 Navigate to the P2 Con parameter 4 Press ENTER Press MODE to select S 232 or S USb...

Page 16: ...the ZERO ENTER key is pressed Weight on the pan when turning the scale on Excessive weight on the pan when zeroing the scale Improper calibration of the scale Damaged load cell Damaged Electronics Er...

Page 17: ...e 1200g 15 Weight Codes and Conversion DISPLAY DESCRIPTION CONVERSION TO GRAMS Ct Net carat 0 2g Lb Lb 453 59237g Oz Ounce 28 349523125g Dr Dram 1 7718451g Ozt Troy ounce 31 1034768g Dwt Penny weight...

Page 18: ...NOTES...

Page 19: ...NOTES...

Page 20: ...Triner Scale Manufacturing Co Inc 8411 Hacks Cross Road Olive Branch MS 38654 800 238 0152 www trinerscale com...

Reviews: