User’s Guide | 1200g

                                

 

 

 7 

 
VIEW AUDIT & CALIBRATION COUNTS

 

Calibration Counter 
Press P key during self-checking, display will show 
number of times scale has been calibrated. When 
the counter becomes full, it will start over  
at 0. 

Audit Operation Counter 
Press S key during self-checking, display will show 
number of audit operations. 

Adjusting the Parameters

 

To access the parameter settings, press the MODE 
key during the start up self-checking. 

F1 Unt will display. Press the MODE key to move 
through the available parameters.  

To access the settings in the parameter showing in 
the display (such as F1 Unt), press the ENTER key. 
The first setting will display (such as “on ct” in the 
F1 Unt parameter). 

To navigate through the available options for the 
setting, press the MODE key (such as “on ct” or oFF 
ct”).  

To accept the option being displayed, press the 
ENTER key. If more than one setting is available in 
the parameter, the next setting will display.  

To save the settings, press the Enter key as needed 
to return to the parameter title (such as F1 Unt), 
then press the ESC key. 

See the following pages for parameters charts. 

Summary of Contents for 1200g High Precision Scale

Page 1: ...UG Version 1 2_12 30 13 1200g High Precision Scale User s Guide 1200g High Precision Scale Legal for Trade Includes RS 232 interface port...

Page 2: ...ross Road Olive Branch MS 38654 800 238 0152 www trinerscale com 2013 Triner Scale All rights reserved CALIBRATION WARNING Calibration AND inspection of calibration properties is prohibited unless don...

Page 3: ...10 Output of the Weight 11 Output to a Receipt Printer 11 Output to a PC 12 Output to a Remote Display 13 Error Codes 14 Weighing Unit Conversions 15 NOTICE The CZ 1200 High Precision scale is a very...

Page 4: ...ting features include automatic zero tracking and an accumulation facility that allows the count to be stored and recalled as an accumulated total Specifications Capacity 1200g x 0 1g NTEP Certified L...

Page 5: ...rs of electricity such as welding equipment or large motors Avoid high humidity that might cause condensation Avoid direct contact with water Do not spray or immerse the scales in water Avoid air move...

Page 6: ...erface ESC Return to normal operation when the scale is in a parameter setting mode avoids saving recently entered changes UNIT MODE UNIT Selects either grams pounds carat ounce etc for the weighing u...

Page 7: ...account for minor drifting or accumulation of material on the platform When the weight goes over 20 of the full capacity automatic zeroing is disabled TARING If necessary zero the scale by pressing th...

Page 8: ...form 2 After the stable icon displays press the PRINT key 3 ACC 1 will display 4 Remove all weight from scale platform then place next item to be weighed on the platform 5 Per 2 above press the PRINT...

Page 9: ...isplay Press the MODE key to move through the available parameters To access the settings in the parameter showing in the display such as F1 Unt press the ENTER key The first setting will display such...

Page 10: ...dwt F2 bl BACKLIGHT SETTING EL on EL backlight always on EL Au EL backlight automatically turn on EL off El backlight always off F3 cal CAUTION Calibration is to be performed by a qualified scale tech...

Page 11: ...n is to be performed by a qualified scale technician ONLY LINEAR CALIBRATION Remove all weight from the pan Press ENTER PIN will display Enter the password by pressing the keys PRINT S P Load 0 will d...

Page 12: ...can select from 600 1200 2400 4800 9600bps SELECT PRINTER OUTPUT MODE If you select P prt or P auto for printer after setting baud rate the display will show tp or LP 50 This is printer mode you can...

Page 13: ...onnect the scale to the printer using a standard DB9 null modem cable 2 Access the Technical Parameters of the CZ 1200 scale see the previous section Adjusting the Parameters 3 Navigate to the P2 Con...

Page 14: ...he CZ 1200 scale see the previous section Adjusting the Parameters 3 Navigate to the P2 Con parameter 4 Press ENTER Press MODE to select S 232 or S USb as required based on the connection 5 Press ENTE...

Page 15: ...dem cable 2 Access the Technical Parameters of the CZ 1200 scale see the previous section Adjusting the Parameters 3 Navigate to the P2 Con parameter 4 Press ENTER Press MODE to select S 232 or S USb...

Page 16: ...the ZERO ENTER key is pressed Weight on the pan when turning the scale on Excessive weight on the pan when zeroing the scale Improper calibration of the scale Damaged load cell Damaged Electronics Er...

Page 17: ...e 1200g 15 Weight Codes and Conversion DISPLAY DESCRIPTION CONVERSION TO GRAMS Ct Net carat 0 2g Lb Lb 453 59237g Oz Ounce 28 349523125g Dr Dram 1 7718451g Ozt Troy ounce 31 1034768g Dwt Penny weight...

Page 18: ...NOTES...

Page 19: ...NOTES...

Page 20: ...Triner Scale Manufacturing Co Inc 8411 Hacks Cross Road Olive Branch MS 38654 800 238 0152 www trinerscale com...

Reviews: