ERROR MESSAGES
OVERLOAD
- If the Micromotor surpasses or exceeds its maximum safety threshold this
message will appear on the display. Too much force on the Micromotor may be the cause.
Try easing the pressure being applied.
ERROR!!!
- This message will appear if the OVERLOAD message is on the display and
extreme force is still being applied. A safety sensor will signal the ERROR!!! Message and
automatically reduce the power supplied to the Micromotor.
BLOCKED
- This message will appear when the motor has been obstructed or prevented
from turning for a prolonged period of time try releasing the Foot Pedal and depressing it
again.
MAINTENANCE
The iON is equipped with high-speed, pre-lubricated bearings.
DO NOT LUBRICATE
THESE BEARINGS.
Periodically check the air ducts at the rear of the Micromotor. Air should always exit from
them. Remove any material that may be obstructing these ducts.
Periodically remove the dust protection cap from the front of the Micromotor and clean it
with compressed air. A soft brush may be used to clean the area around the shaft.
If the iON is not being used for a long period of time, it should be stored in a dry area and
covered to protect it from dust and humidity.