Communication
TCD Electronics | Function Manual | Version: TCD_PROFI_S_EN_200326_E003
10
3.4.
Changing Profibus termination at the transmitter cassette
To change the termination of the Profibus, it is required to remove the transmitter module out of the
transmitter. This can be done following below instructions.
1.
Isolate the device from power.
2.
Unscrew the locking screw of the display and remove the display lid using a wrench bar if
necessary.
3.
Carefully remove display and cable plug.
NOTE:
Use a screwdriver to release the display as shown in photo below. The display is held in place with three
clips.
4.
Remove the screw holding the transmitter cassette. Pull out the transmitter cassette.
5.
Changing termination at the transmitter cassette, by changing the position of the dip switches as
below.
6.
All off: no termination
7.
All on: termination activated