![Trico DR-7 User Manual Download Page 18](http://html.mh-extra.com/html/trico/dr-7/dr-7_user-manual_1162111018.webp)
Page 18 of 28
7.4 – CALIBRATION Tab
The DR-7 has two calibration features OPTICS calibration and END OF OIL sensor calibration located in
the CALIBRATION tab. This helps maintain optimum performance during operation. This is controlled
by the lab technician when maintenance and cleaning is performed or performed on a routine basis before
sample testing. These features are in place to allow the equipment to be in service for lengthy periods of
time without the need to ship the unit back to the factory for optics cleaning or optics calibration.
7.4.1 –
OPTICAL Calibration
The optics guide block is aligned with the LED light source during factory alignment calibration. Unlike
previous DR versions, the optics block should never be removed for cleaning (see section 9 Maintaining
the DR-7). The new design of the Optics block uses a special fluid filled calibration tube for both the
alignment procedure during the factory build, and in calibrating the LED light source by the lab
technician.
The LED light source calibration should be done daily, or every time the instrument is turned on. This is
done by selecting the OPTICS button in the CALIBRATION tab. In the process, the fluid filled
calibration tube is used to focus light changes in the instrument. In addition, an
OPTICAL CALIBRATION
CHECK
is used to check the optical calibration to ensure the unit falls within a specified range. This
check should be performed monthly if the unit is used frequently. Although increased sensitivity has been
enhanced throughout the development of this instrument, because of particle distribution and other
variables within the oil sample being tested, the DR-7 is still considered a trending instrument.