7.
Operating Procedures
Always use the DR-7 on a sturdy, level surface in a well-ventilated room. To start the DR-7 system, flip
the rocker switch on the back of the unit to the “ON” position. The operating system will automatically
start booting and the Trico logo will illuminate. Once the windows system has booted the DR-7 program
will automatically open to the RUN DR tab, otherwise there is a shortcut on the desktop to launch the
DR-7 program.
At system startup the main DR-7 menu appears. The first start up screen is shown below. This start up
screen allows the user to quickly change settings, calibrate optics or end of oil sensor or start the testing
process. At the very top of the main menu six tabs appear:
RUN DR, CHARTING, BASELINE,
CALIBRATION, HELP
, and
ABOUT
, which are explained in further detail.
7.1 – RUN DR Tab
The Run DR tab is the functional work areas for operating the DR-7 system and allows users to change
settings on the fly, or operate the program and circulate the pump.
Located on the right side of the RUN DR tab is the Submenu Selection Area. In this area all sub menues
will appear when the MENU, STANDARD, or OPAQUE buttons are selected.
In the left half of the screen two black blocks appear indicating the DL and DS numbers during the test.
Above these blocks, system messages will appear prompting the user or indicating warning messages
when parameters are out of range.
Three touch buttons are also located at the lower left side of the screen under the DL and DS reading
boxes. These buttons allow the user to start tests using the STANDARD or OPAQUE functions. By
pressing these buttons, a new menu appears in the submenu selection area for running the specific test.
The PUMP button, when pressed rotates the peristaltic pump one revolution and can be used to start the
sample flow through the precipitator tube if the sample viscosity is too high and flow is unable to be
achieved through the current rotation settings. The PUMP button should only be used to start the flow and
should not be continually used to process the sample at a faster pace. Doing so will give false DL and DS
readings because of inconsistency of flow over the magnet assembly.
Submenu Area