Tips:
1. There are two users in total. Each user has 60 records.
2. You can press “ ” button to stop measuring during the
process of the measurement at any time.
3. If the measurement result is out of the measurement range
(SYS: 60mmHg to 230mmHg; or DIA: 40mmHg to 130mmHg;or
Pulse: 40-199 pulse/minute), the LCD will display “out”.
Note:
(1). The symbol “ ” will disappear after successful data transmission, and you may check
your personal health data stored in your mobile device.
(2). If the data transmission fails, the symbol “ ” will remain. The pending measurement
data will be transmitted to your mobile device when next measurement is complete.
(3). If you have set the [Positioning Symbol] on, when you start the measurement, the blood
pressure monitor will detect the position first. The wrist must beat the angle between 30°
and 45°. If it’s out of this angle, the LCD won’t start any measurements and will display
blinking symbol “ ” until you have the correct position. If you get the correct position,
it will display “ ” for about 3 seconds, and then finish the whole measurement.
During the measurement or when the
measurement is finished, press “ ”
button to turn off the monitor. Otherwise it will
power off in one minute.
3.
CAUTION
1.
2.
Interference may occur in the vicinity of equipment marked with the following symbol
. And TMB-1598-BS may interfering vicinity electrical equipment.
Sensitive people, including pregnant women pre-eclamptic and those who implanted
medical electronic instruments, should avoid using the unit whenever possible.
Keep the monitor at least 20 centimeters away from the human body (especially the
head) when the data transmission is proceeding after measurement.
To enable the data transmission function, this product should be paired to Bluetooth
end at 2.4 GHz.
How to mitigate possible interference?
The range between the device and BT end should be reasonably close, from 1 meter
to 10 meters. Please ensure no obstacles between the device and BT end so as to
obtain quality connection and to lower the RF output range.
To avoid interference, other electronic devices (particularly those with wireless
transmission / Transmitter) should be kept at least 1 meter away from the monitor.
1
.
2
.
The
corresponding
time is 22:36.
The
corresponding
date is 8th
May.
Press “ ” button or “ ”
button to get the record you want.
When the monitor is off,
press “ ” button to show
the average value of the
latest three measurement
records. If the records are
less than 3 groups, it will
display the latest record
instead.
The record number, time and
date of the measurement records
will display alternately.
The record
number is
No.1. Six
records in
total.
Recall the Records
MEASUREMENT
16
17
DATA MANAGEMENT