CAUTION

* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.

*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied 

around simultaneously, because this could cause temporary loss of function of those 

simultaneously-used monitoring ME equipment.

*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, 

open the cuff immediately. Prolonged high pressure (cuff pressure 

 300mmHg or constant 

pressure 

 15mmHg for more than 3 minutes) applied to the wrist may lead to an 

ecchymosis.

*Please check that operation of the device does not result in prolonged impairment of patient 

blood circulation.

* When measurement, please avoid compression or restriction of the connection tubing.

* The device cannot be used with HF surgical equipment at the same time.

* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was 

clinically investigated according to the requirements of ISO 81060-2:2013.

* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the 

manufacturer.

* This device is contraindicated for any female who may be suspected of, or is pregnant. 

Besides providing inaccurate readings, the effects of this device on the fetus are unknown.

* Too frequent and consecutive measurements could cause disturbances in blood circulation 

and injuries.

* This unit is not suitable for continuous monitoring during medical emergencies or 

operations.Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen and 

even purple due to a lack of blood.

* When not in use, store the device in a dry room and protect it against extreme moisture, 

heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.

* This device may be used only for the purpose described in this booklet. The manufacturer 

cannot be held liable for damage caused by incorrect application.

*This device comprises sensitive components and must be treated with caution. Observe the 

storage and operating conditions described in this booklet.

* The maximum temperature that the applied part can be achieved is 42.5

 while the 

environmental temperature is 40

.

* The equipment is not AP/APG equipment and not suitable for use in the presence of a 

flammable anesthetic mixture with air of with oxygen or nitrous oxide.

* Warning: No servicing/maintenance while the ME equipment is in use.

* The patient is an intended operator.

* The patient can measure, transmit data and change batteries under normal circumstances 

and maintain the device and its accessories according to the user manual.

* To avoid measurement errors, please avoid the condition of strong electromagnetic field 

radiated interference signal or electrical fast transient/burst signal.

* The blood pressure monitor and the cuff are suitable for use within the patient environment. 

If you are allergic to polyester, nylon or plastic, please don't use this device. 

* During use, the patient will be in contact with the cuff. The materials of the cuff have been 

tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. 

It will not cause any potential sensization or irritation reaction.

* If you experience discomfort during a measurement, such as pain in the wrist or other 

complaints, press the START/STOP button to release the air immediately from the cuff. 

Loosen the cuff and remove it from your wrist.

* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should 

the cuff not deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the 

wrist and press the START/STOP button to stop inflation.

* Before use, make sure the device functions safely and is in proper working condition. 

Check the device, do not use the device if it is damaged in any way. The continuous use of a 

damaged unit may cause injury, improper results, or serious danger.

* Do not wash the cuff in a washing machine or dishwasher!

* The service life of the cuff may vary by the frequency of washing, skin condition, and storage 

state. The typical service life is 10000 times.

* It is recommended that the performance should be checked every 2 years and after 

maintenance and repair, by retesting at least the requirements in limits of the error of the cuff 

pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).

* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to 

the local guidelines.

* Manufacturer will make available on request circuit diagrams, component part lists, 

descriptions, calibration instructions,etc., to assist to service personnel in parts repair.

* The operator shall not touch output of batteries and the patient simultaneously.

* Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth 

to clean the whole unit before and after use. Don’t use any abrasive or volatile cleaners.

* The device doesn’t need to be calibrated within two years of reliable service.

* If you have any problems with this device, such as setting up, maintaining or using, please 

contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in 

the event of malfunctions. The device must only be serviced, repaired and opened by 

individuals at authorized sales/service centers.

* Please report to Transtek if any unexpected operation or events occur.

* Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing 

of small parts. It is dangerous or even fatal.

* Be careful to strangulation due to cables and hoses, particularly due to excessive length.

* At least 30 min required for ME equipment to warm from the minimum storage temperature 

between uses until it is ready for intended use. At least 30 min required for ME equipment to 

cool from the maximum storage temperature between uses until it is ready for intended use.

* This equipment needs to be installed and put into service in accordance with the information 

provided in the ACCOMPANYING DOCUMENTS;

* Wireless communications equipment such as wireless home network devices, mobile 

phones, cordless telephones and their base stations, walkie-talkies can affect this equipment 

and should be kept at least a distance d away from the equipment. The distance d is 

caculated by the MANUFACTURER from the 80 MHz to 5.8 GHz column of Table 4 and Table 

9 of IEC 60601-1-2:2014, as appropriate.

* Please use ACCESSORIES and detachable partes specified/ authorised by MANUFAC-

TURE. Otherwise, it may cause damage to the unit or danger to the user/patients.

* There is no luer lock connectors are used in the construction of tubing, there is a possibility 

that they might be inadvertently connected to intravascular fluid systems, allowing air to be 

pumped into a blood vessel.

* Please use the device under the environment which was provided in the user manual. 

Otherwise, the performance and lifetime of the device will be impacted and reduced.

CAUTION

INTRODUCTION

INTRODUCTION

Summary of Contents for TMB-1598-BS

Page 1: ...age of this product and keep the manual well for your further reference in case you have problems Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB 1598 BS Wrist Type Version 1 0 G...

Page 2: ...blood pressure at home compared to the hospital Is the result the same if measuring on the right wrist TROUBLESHOOTING 24 SPECIFICATIONS 25 CONTACT INFORMATION 26 Safety Information Start Measurement...

Page 3: ...devices patients with pre eclampsia premature ventricular beats atrial fibrillation peripheral arterial disease and patients undergoing intravascular therapy or arterio venous shunt or people who rec...

Page 4: ...and remove it from your wrist If the cuff pressure reaches 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when pressures reaches 40 kPa 300 mmHg detach the cuff from t...

Page 5: ...he average value of blood pressure User 1 Indicate the blood pressure level Blood pressure level indicator Start measurement and save the results for user 1 User 2 Start measurement and save the resul...

Page 6: ...idelines The typical service life of the new and unused batteries is 54 measurements for the operation time is 60s 1 2 3 It is important to set the clock before using your blood pressure monitor so th...

Page 7: ...steps 2 and 3 to confirm Positioning symbol on or off 6 Repeat steps 2 and 3 to confirm the measurement unit 7 After confirming the meausrement unit the LCD will display donE then display all the set...

Page 8: ...st cuff around your wrist leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate Sit comfortably with your tested wrist resting on a flat su...

Page 9: ...l display User ID donE and then turn off Then you can start your measurement now 2 1 Adjust the zero LCD display Inflating and measuring Display and save the result The year date and time will dispaly...

Page 10: ...cinity of equipment marked with the following symbol And TMB 1598 BS may interfering vicinity electrical equipment Sensitive people including pregnant women pre eclamptic and those who implanted medic...

Page 11: ...confirm Then press button to check the records of the selected user 3 1 If you did not get the correct measurement you can delete all the results for the selected user by following steps Take user 1 f...

Page 12: ...t Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine Maintenance In order to get the best performance please follow the instructions be...

Page 13: ...ths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down It is ok for both wrists but there will be some different results for different people We su...

Page 14: ...your monitor with your mobile device Please check below items Bluetooth is ON App Collector is ON Both devices are within the transmission distance of Bluetooth Power supply Battery powered mode 2 AA...

Page 15: ...anagement to medical devices EN 1041 2008 Information supplied by the manufacturer of medical devices EN 60601 1 2006 IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements...

Page 16: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Page 17: ...quency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths f...

Page 18: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Reviews: