Technical description

1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE 
with regard to electromagnetic disturbances for the excepted service life. 

2. Guidance and manufacturer’s declaration-electromagnetic emissions and Immunity.

Table 1 

Guidance and manufacturer’s declaration - electromagnetic emissions

RF emissions
CISPR 11 

RF emissions
CISPR 11 

Group 1 

Class B

Class A

Comply

Compliance

Harmonic emissions
IEC 61000-3-2 

Voltage fluctuations / flicker emissions
IEC 61000-3-3

Emissions test

Table 2 

EMC GUIDANCE

EMC GUIDANCE

Guidance and manufacturer’s declaration – electromagnetic Immunity 

±8 kV contact
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

±8 kV contact  
±2 kV, ±4 kV, ±8 kV,  ±15 kV air 

Electrostatic
discharge (ESD)
IEC 61000-4-2 

±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency

±2 kV for power supply lines
NA
100 kHz repetition frequency

±0.5 kV, ±1 kV, ±2 kV common mode

±0.5 kV, ±1 kV, differential mode

N/A

Electrical fast
transient/burst
IEC 61000-4-4

Surge 
IEC61000-4-5

Immunity Test

Compliance level

IEC 60601-1-2  
Test level 

30 A/m 
50 Hz / 60 Hz

30 A/m  
50 Hz / 60 Hz

Power frequency
magnetic field
IEC 61000-4-8 

Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11   

0% UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT ;
25/30 cycles; Single phase: at 0°.
0% UT; 250 / 300 cycle

0% UT; 0,5 cycle. At 0°, 45°, 
90°, 135°,
180°, 225°, 270° and 315°.
0% UT; 1 cycle and 70% UT  ;
25/30 cycles; Single phase: at 0°.
0% UT; 250 / 300 cycle

Conduced RF
IEC61000-4-6

3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80% AM at 1 kHz

3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio 
bands between 0,15 MHz and 
80 MHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80% AM at 1 kHz

Radiated RF 
IEC61000-4-3

The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments.
Warning:  Don’t be near the active HF surgical equipment and the RF shielded room of 
an ME system for magnetic resonance imaging, where the intensity of EM disturbances 
is high.
Warning:  Use of this equipment adjacent to or stacked with other equipment should 
be avoided because it could result in improper operation. If such use is necessary, this 
equipment and the other equipment should be observed to verify that they are operating 
normally.
Warning:  Use of accessories, transducers and cables other than those specified or 
provided by the manufacturer of this equipment could result in increased electromagnetic 
emissions or decreased electromagnetic immunity of this equipment and result in 
improper operation. 
Warning:  Portable RF communications equipment (including peripherals such as antenna 
cables and external antennas) should be used no closer than 30 cm (12 inches) to any 
part of the equipment TMB-2092-G including cables specified by t he manufacturer. 
Otherwise, degradation of the performance of this equipment could result.   

NOTE  U 

T

   is the a.c. mains voltage prior to application of the test level. 

Summary of Contents for TeleCuff Gen 2

Page 1: ...0 Guangdong Transtek Medical Electronics Co Ltd Zone A No 105 Dongli Road Torch Development District 528437 Zhongshan Guangdong China User Manual Pro Cellular Blood Pressure Monitor TMB 2092 G TeleCu...

Page 2: ...c pressure What is the standard blood pressure classi cation Why does my blood pressure uctuate throughout the day Is the result the same if measuring on the righe arm ABOUT BLOOD PRESSURE TROUBLESHOO...

Page 3: ...vice Date of Manufacture Recyclable Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for a variety...

Page 4: ...ency of washing skin condition and storage state The This device is intended for indoor use This device is portable but it is not intended for use during patient transport This device is not suitable...

Page 5: ...nt icon loose when measuring will appear if the cuff is not wrapped or wrapped too Cuff wearing detection icon Indicates the current measurement mode is Triple Mode and which measurement progress it i...

Page 6: ...e the AC adapter authorized by the manufacturer Please unplug the adapter to depart from the using utility power when you nish the measurement 6V 1A Do not use new and used batteries together Do not u...

Page 7: ...w rst Note B turned off If unsuccessful the monitor will stop searching and enter the standby mode This device is equipped with function of Measurement Reminder Contact Customer Support to enable or d...

Page 8: ...ill appear on the LCD Adjust the zero point In ating and measuring Display the measured result When the TRIPLE MODE button is located on OFF the monitor is under the Single Measurement mode When the m...

Page 9: ...start the second measurement after 60 seconds 3 Then it will repeat the steps to nish the third measurement When three measurements are done the LCD will display the average of 3 readings Tip If the m...

Page 10: ...dry place and avoid the sunshine Avoid intense shaking and collisions Using wet cloths to remove dirt Avoid washing the cuff Avoid touching water clean it with a dry cloth in case Avoid dusty and uns...

Page 11: ...e range Please note that only a physician can tell whether your blood pressure value has reached a dangerous point Blood Pressure Category Normal Elevated High Blood Pressure Hypertension Stage 1 High...

Page 12: ...K THIS REMEDY No power Low Battery Error message Display can not light up Batteries are exhausted Replace with new batteries Install the batteries or plug in the adapter properly Replace with new batt...

Page 13: ...ressure 0mmHg 299mmHg SYS 60 mmHg 230 mmHg DIA 40 mmHg 130 mmHg Pulse value 40 199 beat minute Device Classi cation Expected Lifetime Battery Powered Mode Internally Powered ME Equipment AC Adaptor Po...

Page 14: ...45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 0 UT 250 300 cycle Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz...

Page 15: ...TE Band 13 17 Pulse modulation 217 Hz GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation 18 Hz GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz Bluetoot...